The influence of haemolysis in patient samples on biochemical tests analysed using Roche Cobas <sup>®</sup> 8000 analyzer

BackgroundModern analyzers employ the haemolysis index (HI) to identify interference in biochemical assays, yet manufacturer-defined HI thresholds may be inappropriate for true haemolysis effects, resulting in unnecessary sample rejections. This study aimed to validate these thresholds using non-simulated hemolyzed patient samples.MethodsPaired samples (hemolyzed primary and non-hemolyzed recollected) from 678 patients were analysed for haemolysis interference. Biochemical analytes and serum indices were measured using a Roche Cobas<sup>®</sup> 8000 analyzer. Haemolysis effects on test results and lipaemia index (LI) were assessed. HI thresholds were derived from reference change value (RCV) limits and regression of HI versus percentage bias, then compared to the conventional 10% deviation criterion and Roche-defined cut-offs.ResultsSamples exhibited predominantly moderate haemolysis (72.3%, HI: 101-300). Strong HI correlations were observed for lactate dehydrogenase (51% change per 100-unit HI, <i>R</i><sup><i>2</i></sup> = 0.6524, <i>P</i> < .0001), potassium (14% per 100-unit HI, <i>R</i><sup><i>2</i></sup> = 0.5630, <i>P</i> < .0001), and sodium (-0.6% per 100-unit HI, <i>R</i><sup><i>2</i></sup> = 0.5414, <i>P</i> < .0001). Elevated biases exceeded the RCV for these analytes, plus ammonia, aspartate aminotransferase, creatine kinase, γ-glutamyltransferase, and bilirubin-direct, whereas sodium showed a clinically significant reduction at heavy haemolysis (HI 560). RCV-derived thresholds exhibited comparable or higher than 10% change and Roche cut-offs. The elevated LI in hemolyzed samples with HI greater than 100 decreased significantly after recollection.ConclusionsPatient-based haemolysis data indicated that biases for most analytes remain within clinically acceptable limits, suggesting the manufacturer's HI thresholds may overestimate interference, supporting lab-validated, RCV-based cut-offs enhance clinical relevance and decrease unnecessary sample rejection.

Authors

Institutions

Publication Details

Journal
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Published
2025-06-24
DOI
https://doi.org/10.1177/00045632251356827
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The influence of haemolysis in patient samples on biochemical tests analysed using Roche Cobas ® 8000 analyzer

Ying‐Chun Lin, Shu‐Chu Shiesh, Yu-Chen Hung, Chun‐An Cheng et al.
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
article

The influence of haemolysis in patient samples on biochemical tests analysed using Roche Cobas ® 8000 analyzer

Ying‐Chun Lin, Shu‐Chu Shiesh, Yu-Chen Hung, Chun‐An Cheng, Wei-Ling Lin, Yu Wei Chiu
article en

Abstract

BackgroundModern analyzers employ the haemolysis index (HI) to identify interference in biochemical assays, yet manufacturer-defined HI thresholds may be inappropriate for true haemolysis effects, resulting in unnecessary sample rejections. This study aimed to validate these thresholds using non-simulated hemolyzed patient samples.MethodsPaired samples (hemolyzed primary and non-hemolyzed recollected) from 678 patients were analysed for haemolysis interference. Biochemical analytes and serum indices were measured using a Roche Cobas<sup>®</sup> 8000 analyzer. Haemolysis effects on test results and lipaemia index (LI) were assessed. HI thresholds were derived from reference change value (RCV) limits and regression of HI versus percentage bias, then compared to the conventional 10% deviation criterion and Roche-defined cut-offs.ResultsSamples exhibited predominantly moderate haemolysis (72.3%, HI: 101-300). Strong HI correlations were observed for lactate dehydrogenase (51% change per 100-unit HI, <i>R</i><sup><i>2</i></sup> = 0.6524, <i>P</i> < .0001), potassium (14% per 100-unit HI, <i>R</i><sup><i>2</i></sup> = 0.5630, <i>P</i> < .0001), and sodium (-0.6% per 100-unit HI, <i>R</i><sup><i>2</i></sup> = 0.5414, <i>P</i> < .0001). Elevated biases exceeded the RCV for these analytes, plus ammonia, aspartate aminotransferase, creatine kinase, γ-glutamyltransferase, and bilirubin-direct, whereas sodium showed a clinically significant reduction at heavy haemolysis (HI 560). RCV-derived thresholds exhibited comparable or higher than 10% change and Roche cut-offs. The elevated LI in hemolyzed samples with HI greater than 100 decreased significantly after recollection.ConclusionsPatient-based haemolysis data indicated that biases for most analytes remain within clinically acceptable limits, suggesting the manufacturer's HI thresholds may overestimate interference, supporting lab-validated, RCV-based cut-offs enhance clinical relevance and decrease unnecessary sample rejection.

Annals of Clinical Biochemistry International Journal of Laboratory MedicineVol. 63(1)
Purdue University West Lafayette (US), Tainan Municipal Hospital (TW), National Cheng Kung University (TW)
National Cheng Kung University Hospital
Openalex Percentile: Top 81%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.