Investigation of the Optimal Parameters for Block Sequential Regularized Expectation Maximization (BSREM) Reconstruction in Bone SPECT Using a Ring-type Semiconductor SPECT/CT
As reconstruction parameters for BSREM reconstruction in bone SPECT on a ring-type semiconductor SPECT/CT system, a γ value of 5.0 and a β value of 0.9 were considered suitable for image quality and quantitative evaluation.
Authors
- Akihito Usui (ORCID: https://orcid.org/0000-0003-1889-7496)
- Kentaro Takanami (ORCID: https://orcid.org/0000-0002-0098-7760)
- Hiroyasu Kodama
- Tatsuo Takeuchi (ORCID: https://orcid.org/0009-0008-8786-515X)
- Tomohiro Kaneta (ORCID: https://orcid.org/0000-0003-1146-5092)
- Yoshitaka Tanaka (ORCID: https://orcid.org/0000-0001-8083-845X)