Functional Outcomes and Patient Satisfaction With a Four-Site Myoelectric Prosthesis System Following the Starfish Procedure: A Case Series

ABSTRACT Introduction Partial hand amputations significantly impair fine motor skills and daily functionality. Myoelectric prostheses are a functional solution but are limited by suboptimal signal transmission in the hand and a small number of sites for control. The starfish procedure enhances the interface between residual limb muscles and prosthetic sensors for improved control of prosthetic digits. Methods This prospective case series evaluated functional outcomes and patient satisfaction in individuals with partial hand amputation who had previously undergone the starfish procedure and were then fitted with a commercially available four-site myoelectric prosthesis. Assessments included the Southampton Hand Assessment Procedure (SHAP), clothespin relocation test (CRT), NASA Task Load Index, and a Digit Control Accuracy Test (DCAT), along with patient satisfaction surveys. Three patients, each with a four-finger amputation, were assessed with their original prosthesis at baseline and with the research prosthesis during follow-up sessions. Results The median follow-up time was 13 months. Following an initial adaptation phase with the research prosthesis, patients showed significant improvements in SHAP Index of Function, with a median increase of 22 points, and CRT times, with a median decrease of 1.8 seconds for down trials and 0.9 seconds for up trials, indicating enhanced hand function and dexterity compared to their original prosthesis. The NASA Task Load Index revealed a median workload decrease of −2, suggesting intuitive control with the research prosthesis, despite individual variations. The DCAT confirmed high success rates in intended movements, demonstrating a learning curve with increased proficiency over time and ability for individualized digit control over. Patient satisfaction was positive, supporting the clinical efficacy of the four-site prosthetic system. Discussion The integration of a four-site myoelectric prosthesis in patients who had undergone the starfish procedure allowed for independent digit control and demonstrated improved outcomes compared to their original prostheses. Independent digit control enables more precise object manipulation and naturalistic hand movements, potentially improving performance of activities requiring fine motor control such as typing, fastening buttons, and handling small objects. While the study faced limitations like small sample size and follow-up variability due to COVID-19 restrictions, the results indicate a clear benefit, advocating for further research into this prosthetic approach. Future studies should aim for a broader participant base, extended follow-up, and comparisons with no prosthesis and passive devices to solidify these findings. Conclusions Results from this preliminary investigation suggest that advanced myoelectric prostheses enabling individual digit control may enhance functionality and user experience in patients who have previously undergone the starfish procedure. The observed improvements in task performance and device interaction, while limited by the small cohort size and variable follow-up period, support expanding availability of these functionally enhancing technologies to appropriate candidates. Clinical Relevance The use of a four-site myoelectric prosthesis in patients who have undergone the starfish procedure shows significant potential for improving outcomes for individuals with partial hand amputation, justifying further research to fully understand its benefits compared to other treatment options.

Authors

Publication Details

Journal
JPO Journal of Prosthetics and Orthotics
Published
2025-06-22
DOI
https://doi.org/10.1097/jpo.0000000000000558
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Functional Outcomes and Patient Satisfaction With a Four-Site Myoelectric Prosthesis System Following the Starfish Procedure: A Case Series

Kennedy Gachigi, Raymond Glenn Gaston, Bryan J. Loeffler, Daire McGuinness et al.
JPO Journal of Prosthetics and Orthotics
Muscle activation and electromyography studies
article

Functional Outcomes and Patient Satisfaction With a Four-Site Myoelectric Prosthesis System Following the Starfish Procedure: A Case Series

Kennedy Gachigi, Raymond Glenn Gaston, Bryan J. Loeffler, Daire McGuinness, Brian Monroe, Phil Stevens, James Tyler Frix, Ryan Serbin, Amy Todd, Lynsay Emmrich
article en

Abstract

ABSTRACT Introduction Partial hand amputations significantly impair fine motor skills and daily functionality. Myoelectric prostheses are a functional solution but are limited by suboptimal signal transmission in the hand and a small number of sites for control. The starfish procedure enhances the interface between residual limb muscles and prosthetic sensors for improved control of prosthetic digits. Methods This prospective case series evaluated functional outcomes and patient satisfaction in individuals with partial hand amputation who had previously undergone the starfish procedure and were then fitted with a commercially available four-site myoelectric prosthesis. Assessments included the Southampton Hand Assessment Procedure (SHAP), clothespin relocation test (CRT), NASA Task Load Index, and a Digit Control Accuracy Test (DCAT), along with patient satisfaction surveys. Three patients, each with a four-finger amputation, were assessed with their original prosthesis at baseline and with the research prosthesis during follow-up sessions. Results The median follow-up time was 13 months. Following an initial adaptation phase with the research prosthesis, patients showed significant improvements in SHAP Index of Function, with a median increase of 22 points, and CRT times, with a median decrease of 1.8 seconds for down trials and 0.9 seconds for up trials, indicating enhanced hand function and dexterity compared to their original prosthesis. The NASA Task Load Index revealed a median workload decrease of −2, suggesting intuitive control with the research prosthesis, despite individual variations. The DCAT confirmed high success rates in intended movements, demonstrating a learning curve with increased proficiency over time and ability for individualized digit control over. Patient satisfaction was positive, supporting the clinical efficacy of the four-site prosthetic system. Discussion The integration of a four-site myoelectric prosthesis in patients who had undergone the starfish procedure allowed for independent digit control and demonstrated improved outcomes compared to their original prostheses. Independent digit control enables more precise object manipulation and naturalistic hand movements, potentially improving performance of activities requiring fine motor control such as typing, fastening buttons, and handling small objects. While the study faced limitations like small sample size and follow-up variability due to COVID-19 restrictions, the results indicate a clear benefit, advocating for further research into this prosthetic approach. Future studies should aim for a broader participant base, extended follow-up, and comparisons with no prosthesis and passive devices to solidify these findings. Conclusions Results from this preliminary investigation suggest that advanced myoelectric prostheses enabling individual digit control may enhance functionality and user experience in patients who have previously undergone the starfish procedure. The observed improvements in task performance and device interaction, while limited by the small cohort size and variable follow-up period, support expanding availability of these functionally enhancing technologies to appropriate candidates. Clinical Relevance The use of a four-site myoelectric prosthesis in patients who have undergone the starfish procedure shows significant potential for improving outcomes for individuals with partial hand amputation, justifying further research to fully understand its benefits compared to other treatment options.

JPO Journal of Prosthetics and OrthoticsVol. 37(4)
Quality Education
Openalex Percentile: Top 87%
Muscle activation and electromyography studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.