Evaluation of the Accuracy of an Endomotor-Integrated Apex Locator Versus a Standalone Electronic Apex Locator in Teeth with Simulated Apical Root Resorption: An In Vitro Study

Objective: This in vitro study aimed to evaluate and compare the accuracy of an endomotor-integrated apex locator system and a conventional electronic apex locator in determining the working length (WL) of root canals with simulated apical root resorption, in both straight and curved canal anatomies. Methods: Sixty-seven extracted mandibular molars with either straight distal canals or curved mesiobuccal canals were included. Simulated external apical resorption was created using a standardized protocol. Actual working length was determined under a stereomicroscope. WL was then measured using an integrated system (EndoRadar Pro) and an electronic apex locator (Dentaport ZX) shaping procedures. The differences between actual and measured WL values were calculated. Data were analyzed using the Kolmogorov-Smirnov test and Chi-square test with a ±0.5 mm tolerance range. Results: All tested devices demonstrated clinically acceptable accuracy within the ±0.5 mm range. In curved canals, the integrated system used with ProGlider showed higher accuracy compared to the systems used with PTN X1 and Dentaport ZX; however, this difference was not statistically significant (Pearson Chi-Square, p = 0.128). In straight canals no statistically significant differences were observed (Pearson Chi-Square, p = 0.155). Conclusions: Both integrated system used with EndoRadar Pro and electronic measurements performed with Dentaport ZX provide reliable measurements for WL determination in the presence of apical root resorption. The integrated system used with ProGlider may offer enhanced accuracy in anatomically complex, curved canals, supporting their clinical use in such challenging scenarios.

Authors

Institutions

Publication Details

Journal
European Journal of Therapeutics
Published
2025-06-20
DOI
https://doi.org/10.58600/eurjther2693
Primary Topic
Endodontics and Root Canal Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Accuracy of an Endomotor-Integrated Apex Locator Versus a Standalone Electronic Apex Locator in Teeth with Simulated Apical Root Resorption: An In Vitro Study

Hatice Harorlı, Simay Koç, Damla Erkal, Yunus Emre Çakmak
European Journal of Therapeutics
Endodontics and Root Canal Treatments
article

Evaluation of the Accuracy of an Endomotor-Integrated Apex Locator Versus a Standalone Electronic Apex Locator in Teeth with Simulated Apical Root Resorption: An In Vitro Study

Hatice Harorlı, Simay Koç, Damla Erkal, Yunus Emre Çakmak
article en

Abstract

Objective: This in vitro study aimed to evaluate and compare the accuracy of an endomotor-integrated apex locator system and a conventional electronic apex locator in determining the working length (WL) of root canals with simulated apical root resorption, in both straight and curved canal anatomies. Methods: Sixty-seven extracted mandibular molars with either straight distal canals or curved mesiobuccal canals were included. Simulated external apical resorption was created using a standardized protocol. Actual working length was determined under a stereomicroscope. WL was then measured using an integrated system (EndoRadar Pro) and an electronic apex locator (Dentaport ZX) shaping procedures. The differences between actual and measured WL values were calculated. Data were analyzed using the Kolmogorov-Smirnov test and Chi-square test with a ±0.5 mm tolerance range. Results: All tested devices demonstrated clinically acceptable accuracy within the ±0.5 mm range. In curved canals, the integrated system used with ProGlider showed higher accuracy compared to the systems used with PTN X1 and Dentaport ZX; however, this difference was not statistically significant (Pearson Chi-Square, p = 0.128). In straight canals no statistically significant differences were observed (Pearson Chi-Square, p = 0.155). Conclusions: Both integrated system used with EndoRadar Pro and electronic measurements performed with Dentaport ZX provide reliable measurements for WL determination in the presence of apical root resorption. The integrated system used with ProGlider may offer enhanced accuracy in anatomically complex, curved canals, supporting their clinical use in such challenging scenarios.

European Journal of Therapeutics
Akdeniz University (TR), Burdur Mehmet Akif Ersoy Üniversitesi (TR)
Openalex Percentile: Top 80%
Endodontics and Root Canal Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.