Multiple Shear-Shift Absolute Calibration of Reference Surface for White-Light Interferometers

With advantages for testing mid-frequency form error and strong curved surfaces, white-light interferometers (WLIs) have been widely used for testing advanced X-ray surfaces with sub-nanometer accuracy. Calibration of the reference error is the prerequisite. The commonly utilized random averaging method requires a flat with high form accuracy and the information of the correlation length of the test flat which can hardly be known ahead. For calibrating the reference surface with the accuracy required by most advanced X-ray surfaces, a multiple shear-shift (MSS) absolute test method for WLIs is proposed. The MSS method is conducted by performing multiple sets of measurements on a test surface. Each set involves one measurement at the original position and two measurements with the test surface translated along the two orthogonal directions with different distances of integer pixels, respectively. The mathematical model of reference retrieval is known, however the result is sensitive to environment disturbance and shear-shift errors. The influence of these source of errors is evaluated and optimal acquisition parameters are determined. The reference surface of a 1× objective lens with 8.6mm square view field was calibrated. We could achieve the repeatability of PV 0.18 nm and RMS 0.03 nm with a normal accuracy test surface and the optimal acquisition parameters. The accuracy is PV 0.23 nm and RMS 0.03 nm by comparing the calibration results that obtained by random averaging method utilizing a flat with extreme form accuracy. By comparisons, the repeatability and accuracy drop to unaccepted level with inappropriate acquisition parameters. The proposed method offers a new reference surface calibration strategy with determined model for guaranteeing the absolute measurement accuracy of X-ray surfaces utilizing WLIs.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Instrumentation and Measurement
Published
2025-01-01
DOI
https://doi.org/10.1109/tim.2025.3581630
Citations
3
Primary Topic
Optical measurement and interference techniques
Type
article
Field-Weighted Citation Impact
3.83

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multiple Shear-Shift Absolute Calibration of Reference Surface for White-Light Interferometers

Antong Huang, Shuai Xue, Xibo Gao, Yunfeng Mao et al.
3 citations
IEEE Transactions on Instrumentation and Measurement
Optical measurement and interference techniques
3.83
article

Multiple Shear-Shift Absolute Calibration of Reference Surface for White-Light Interferometers

Antong Huang, Shuai Xue, Xibo Gao, Yunfeng Mao, Zubo Hu
article en
3 citations

Abstract

With advantages for testing mid-frequency form error and strong curved surfaces, white-light interferometers (WLIs) have been widely used for testing advanced X-ray surfaces with sub-nanometer accuracy. Calibration of the reference error is the prerequisite. The commonly utilized random averaging method requires a flat with high form accuracy and the information of the correlation length of the test flat which can hardly be known ahead. For calibrating the reference surface with the accuracy required by most advanced X-ray surfaces, a multiple shear-shift (MSS) absolute test method for WLIs is proposed. The MSS method is conducted by performing multiple sets of measurements on a test surface. Each set involves one measurement at the original position and two measurements with the test surface translated along the two orthogonal directions with different distances of integer pixels, respectively. The mathematical model of reference retrieval is known, however the result is sensitive to environment disturbance and shear-shift errors. The influence of these source of errors is evaluated and optimal acquisition parameters are determined. The reference surface of a 1× objective lens with 8.6mm square view field was calibrated. We could achieve the repeatability of PV 0.18 nm and RMS 0.03 nm with a normal accuracy test surface and the optimal acquisition parameters. The accuracy is PV 0.23 nm and RMS 0.03 nm by comparing the calibration results that obtained by random averaging method utilizing a flat with extreme form accuracy. By comparisons, the repeatability and accuracy drop to unaccepted level with inappropriate acquisition parameters. The proposed method offers a new reference surface calibration strategy with determined model for guaranteeing the absolute measurement accuracy of X-ray surfaces utilizing WLIs.

IEEE Transactions on Instrumentation and MeasurementVol. 74
National University of Defense Technology (CN)
National Natural Science Foundation of China, Natural Science Foundation of Hunan Province
Climate action
Openalex Percentile: Top 6%
Optical measurement and interference techniques
3.83
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.