Discharge Correction Under Asymmetric Electrode-Electrolyte Interfaces in Bipolar Stimulation

Neural electrical stimulators have been widely used in biomedical applications, especially those with multiple channels. However, the electrode-electrolyte interfaces are usually not identical, which is rarely discussed in stimulator design. This paper first analyzes the stimulation and discharge phases in bipolar stimulation when the interfaces are not identical, proving that the asymmetry will lead to incomplete discharge. To address this issue, we propose a tripolar stimulator by introducing a return electrode to discharge the working and counter electrodes separately. The prototype of a 4-channel tripolar stimulator is designed and fabricated in a 180-nm CMOS technology, with each stimulation channel occupying 0.176 mm2. The maximum stimulation current is 2.8 mA. After discharge, residual voltages are near-zero in both electrical and in vitro tests.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Circuits & Systems II Express Briefs
Published
2025-06-18
DOI
https://doi.org/10.1109/tcsii.2025.3580714
Citations
1
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
0.91

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Discharge Correction Under Asymmetric Electrode-Electrolyte Interfaces in Bipolar Stimulation

Yixin Zhou, Keping Wang, Jinyan Wu, Simeng Yin
1 citations
IEEE Transactions on Circuits & Systems II Express Briefs
Neuroscience and Neural Engineering
0.91
article

Discharge Correction Under Asymmetric Electrode-Electrolyte Interfaces in Bipolar Stimulation

Yixin Zhou, Keping Wang, Jinyan Wu, Simeng Yin
article en
1 citations

Abstract

Neural electrical stimulators have been widely used in biomedical applications, especially those with multiple channels. However, the electrode-electrolyte interfaces are usually not identical, which is rarely discussed in stimulator design. This paper first analyzes the stimulation and discharge phases in bipolar stimulation when the interfaces are not identical, proving that the asymmetry will lead to incomplete discharge. To address this issue, we propose a tripolar stimulator by introducing a return electrode to discharge the working and counter electrodes separately. The prototype of a 4-channel tripolar stimulator is designed and fabricated in a 180-nm CMOS technology, with each stimulation channel occupying 0.176 mm2. The maximum stimulation current is 2.8 mA. After discharge, residual voltages are near-zero in both electrical and in vitro tests.

IEEE Transactions on Circuits & Systems II Express BriefsVol. 72(8)
Tianjin University (CN), Southeast University (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 26%
Neuroscience and Neural Engineering
0.91
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Discharge Correction Under Asymmetric Electrode-Electrolyte Interfaces in Bipolar Stimulation — Yixin Zhou, Keping Wang, et al. · IEEE Transactions on Circuits & Systems II Express Briefs (2025) | TGRS Research Map | TGRS