Time-efficient simultaneous fat and water cardiac cine imaging using spiral magnetic resonance imaging
The proposed method to obtain W/F separated cine images provides better temporal efficiency and fewer artifacts compared to conventional Cartesian bSSFP sequences. The 2-TE bSSFP features the highest artifact level, including susceptibility artifacts and fat/water swaps. The proposed method reduces scan time by approximately 50% with similar spatial and temporal resolution with lower specific absorption rate (SAR). The contrast between the blood pool and myocardium is higher when using the spiral readout (p≤0.05). The results suggest that the presented sequence has potential to facilitate simultaneous imaging for water and fat components in a cine scan while shortening exam time and lowering SAR.
Authors
- Tzu Cheng Chao (ORCID: https://orcid.org/0000-0002-5501-9572)
- Dinghui Wang (ORCID: https://orcid.org/0000-0002-7403-3759)
- Tim Leiner (ORCID: https://orcid.org/0000-0003-1885-5499)
- James G. Pipe (ORCID: https://orcid.org/0000-0002-7858-1162)
Institutions
- Mayo Clinic (US)
- University of Wisconsin–Madison (US)
- Mayo Clinic in Arizona (US)
- Mayo Clinic in Florida (US)
Publication Details
- Journal
- Journal of Cardiovascular Magnetic Resonance
- Published
- 2025-01-01
- DOI
- https://doi.org/10.1016/j.jocmr.2025.101926
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00