Quantification of the difference in hounsfield units of an electron density phantom between a conventional and standing computed tomography machine

In human CT imaging, opportunistic calculation of bone mineral density (BMD) is used to screen for osteoporosis, where BMD is estimated by Hounsfield Units (HU). There is potential to use HU for estimation of BMD in veterinary imaging. However, HU values can vary between different machines and over time. We undertook a serial measurements study to determine (1) the extent of the difference of the mean HU and noise (HU standard deviation) of eight regions of interest (ROI) of an electron density phantom scanned on two CT machines, a Siemens CT (conventional) machine and the Equina by Asto CT (standing) machine; and (2) to determine if the HU values obtained on each machine differed over time. The mean HU from 30 scans of an electron density phantom were generated from each of the two CT machines (N = 60; conventional and standing). Differences between machines were assessed using a linear mixed effects model. The mean HU of each positive value ROI from the conventional machine was decreased compared with the standing machine (P < 0.001) and the difference was greatest for the denser phantom inserts. A calibration curve equation utilising the four bone-equivalent insert plugs of the phantom, with known calcium hydroxyapatite (CaHA) concentration, was generated using linear regression to represent CaHA, and therefore BMD, for the conventional machine (CaHA = -29.58 + 0.98xROI<sub>mean</sub>) and the standing machine (CaHA = -54.53 + 0.95xROI<sub>mean</sub>). For opportunistic calculation of BMD from HU in horses, the differences in HU observed between different CT machines warrants calibration of HU using a bone equivalent phantom.

Authors

Institutions

Publication Details

Journal
Veterinary Research Communications
Published
2025-06-17
DOI
https://doi.org/10.1007/s11259-025-10789-7
Citations
3
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantification of the difference in hounsfield units of an electron density phantom between a conventional and standing computed tomography machine

Peter Muir, R. Chris Whitton, Cathy Beck, Peta Lee Hitchens et al.
3 citations
Veterinary Research Communications
Advanced X-ray and CT Imaging
2.02
article

Quantification of the difference in hounsfield units of an electron density phantom between a conventional and standing computed tomography machine

Peter Muir, R. Chris Whitton, Cathy Beck, Peta Lee Hitchens, Ashleigh V. Morrice‐West
article en
3 citations

Abstract

In human CT imaging, opportunistic calculation of bone mineral density (BMD) is used to screen for osteoporosis, where BMD is estimated by Hounsfield Units (HU). There is potential to use HU for estimation of BMD in veterinary imaging. However, HU values can vary between different machines and over time. We undertook a serial measurements study to determine (1) the extent of the difference of the mean HU and noise (HU standard deviation) of eight regions of interest (ROI) of an electron density phantom scanned on two CT machines, a Siemens CT (conventional) machine and the Equina by Asto CT (standing) machine; and (2) to determine if the HU values obtained on each machine differed over time. The mean HU from 30 scans of an electron density phantom were generated from each of the two CT machines (N = 60; conventional and standing). Differences between machines were assessed using a linear mixed effects model. The mean HU of each positive value ROI from the conventional machine was decreased compared with the standing machine (P < 0.001) and the difference was greatest for the denser phantom inserts. A calibration curve equation utilising the four bone-equivalent insert plugs of the phantom, with known calcium hydroxyapatite (CaHA) concentration, was generated using linear regression to represent CaHA, and therefore BMD, for the conventional machine (CaHA = -29.58 + 0.98xROI<sub>mean</sub>) and the standing machine (CaHA = -54.53 + 0.95xROI<sub>mean</sub>). For opportunistic calculation of BMD from HU in horses, the differences in HU observed between different CT machines warrants calibration of HU using a bone equivalent phantom.

Veterinary Research CommunicationsVol. 49(4)
University of Wisconsin–Madison (US), The University of Melbourne (AU)
Racing Victoria, Australian Government, University of Melbourne, State Government of Victoria
Openalex Percentile: Top 14%
Advanced X-ray and CT Imaging
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.