Variable Flip Angle Optimization for Fetal Brain Imaging With Reduced Specific Absorption Rate

Half-Fourier Acquisition with Single-Shot Turbo Spin Echo (HASTE) scans are routinely used for fetal brain imaging, but they have a high specific absorption rate (SAR) and can be inefficient due to SAR limits. Here, we have designed an optimized variable flip angle (VFA) pattern for fetal HASTE neuroimaging to reduce both the SAR and repetition time (TR) of the HASTE sequence while maintaining similar image quality to standard fetal HASTE imaging with a constant flip angle (CFA). The VFA pattern was optimized by minimizing the difference in expected signal between the VFA and CFA scans while constraining the SAR of the VFA scan to no more than 65% of the SAR of the CFA scan and reducing the TR by 29%. The expected signal was calculated using an extended phase graph formalism, and simulations were used to predict the performance of the two scans using different image quality metrics. The proposed VFA and standard CFA scans were tested in phantoms and fetuses at 3 T. SAR and acquisition times were recorded, and image quality was rated by three radiologists. The VFA scan showed similar signal-to-noise ratio and contrast-to-noise ratio values but slightly lower signal and relative contrast values than the CFA scan in phantom studies. In vivo, the VFA scan yielded significantly reduced SAR, measurement times, and total scan times. There was no significant difference in overall image quality ratings between the VFA and CFA scans. An optimized VFA scan can provide 65% of the SAR and 71% of the acquisition time of a CFA scan while being diagnostically equivalent. Lower SAR reduces heating, eliminates SAR pauses, and allows accelerated scans by reducing the TR. The time saved by faster HASTE acquisitions increases patient comfort and may be used to repeat scans with excessive fetal motion or to perform advanced sequences.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2025-06-16
DOI
https://doi.org/10.1002/nbm.70075
Citations
1
Primary Topic
Fetal and Pediatric Neurological Disorders
Type
article
Field-Weighted Citation Impact
3.49

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Variable Flip Angle Optimization for Fetal Brain Imaging With Reduced Specific Absorption Rate

Dominique Franson, Clemente Velasco‐Annis, Edward Yang, Hongli Fan et al.
1 citations
NMR in Biomedicine
Fetal and Pediatric Neurological Disorders
3.49
article

Variable Flip Angle Optimization for Fetal Brain Imaging With Reduced Specific Absorption Rate

Dominique Franson, Clemente Velasco‐Annis, Edward Yang, Hongli Fan, Cesar Alves, Ali Gholipour, Stephen Cauley, Camilo Calixto, Suely Fazio Ferraciolli, Camilo Jaimes
article en
1 citations

Abstract

Half-Fourier Acquisition with Single-Shot Turbo Spin Echo (HASTE) scans are routinely used for fetal brain imaging, but they have a high specific absorption rate (SAR) and can be inefficient due to SAR limits. Here, we have designed an optimized variable flip angle (VFA) pattern for fetal HASTE neuroimaging to reduce both the SAR and repetition time (TR) of the HASTE sequence while maintaining similar image quality to standard fetal HASTE imaging with a constant flip angle (CFA). The VFA pattern was optimized by minimizing the difference in expected signal between the VFA and CFA scans while constraining the SAR of the VFA scan to no more than 65% of the SAR of the CFA scan and reducing the TR by 29%. The expected signal was calculated using an extended phase graph formalism, and simulations were used to predict the performance of the two scans using different image quality metrics. The proposed VFA and standard CFA scans were tested in phantoms and fetuses at 3 T. SAR and acquisition times were recorded, and image quality was rated by three radiologists. The VFA scan showed similar signal-to-noise ratio and contrast-to-noise ratio values but slightly lower signal and relative contrast values than the CFA scan in phantom studies. In vivo, the VFA scan yielded significantly reduced SAR, measurement times, and total scan times. There was no significant difference in overall image quality ratings between the VFA and CFA scans. An optimized VFA scan can provide 65% of the SAR and 71% of the acquisition time of a CFA scan while being diagnostically equivalent. Lower SAR reduces heating, eliminates SAR pauses, and allows accelerated scans by reducing the TR. The time saved by faster HASTE acquisitions increases patient comfort and may be used to repeat scans with excessive fetal motion or to perform advanced sequences.

NMR in BiomedicineVol. 38(7)
Boston Children's Hospital (US), Harvard University (US), University of California, Irvine (US), Massachusetts General Hospital (US), Medical Solutions (GB), Irvine University (US), Siemens Healthcare (United States) (US)
Siemens Medical Solutions USA, National Institutes of Health, NIH Office of the Director, National Institute of Biomedical Imaging and Bioengineering
Openalex Percentile: Top 9%
Fetal and Pediatric Neurological Disorders
3.49
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.