Setting the Emissivity of an Imaging Bolometer in the Surface Temperature Profile Measurement of SiC-Based MEMS Heaters

The proper usage of a bandwidth-limited imaging bolometer for the measurement of the lateral temperature profile of microstructures in Silicon-Carbide (SiC) is analyzed. The SiC spectral emissivity, ϵSiC(λ), has a dip at λ∼12μm, which is in the band of a typical commercially available instrument and complicates the selection of the value of the equivalent emissivity, ϵeq,SiC, in the instrument settings. The impact is analyzed by deduction using simulation, and by experimental validation. Membranes of 3C-SiC of 1000 μm diameter and 3 μm thickness have been fabricated on Si wafers, with integrated poly-SiC resistors for both membrane heating and on-membrane temperature measurement for calibration purposes. The optimum setting was found as ϵeq,SiC = 0.705 ± 0.025 by deduction and as ϵeq,SiC = 0.66 ± 0.06 by experimental validation in the temperature range 120 °C to 400 °C. The apparent temperature coefficient of emissivity, TCE< 2 × 10−4 °C−1 is due to the shift of the Wien peak wavelength relative to the instrument’s sensitivity band.

Authors

Institutions

Publication Details

Journal
Metrology
Published
2025-06-17
DOI
https://doi.org/10.3390/metrology5020036
Primary Topic
Thermal Radiation and Cooling Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Setting the Emissivity of an Imaging Bolometer in the Surface Temperature Profile Measurement of SiC-Based MEMS Heaters

David Bilby, R.F. Wolffenbuttel, Jaco Visser
Metrology
Thermal Radiation and Cooling Technologies
article

Setting the Emissivity of an Imaging Bolometer in the Surface Temperature Profile Measurement of SiC-Based MEMS Heaters

David Bilby, R.F. Wolffenbuttel, Jaco Visser
article en

Abstract

The proper usage of a bandwidth-limited imaging bolometer for the measurement of the lateral temperature profile of microstructures in Silicon-Carbide (SiC) is analyzed. The SiC spectral emissivity, ϵSiC(λ), has a dip at λ∼12μm, which is in the band of a typical commercially available instrument and complicates the selection of the value of the equivalent emissivity, ϵeq,SiC, in the instrument settings. The impact is analyzed by deduction using simulation, and by experimental validation. Membranes of 3C-SiC of 1000 μm diameter and 3 μm thickness have been fabricated on Si wafers, with integrated poly-SiC resistors for both membrane heating and on-membrane temperature measurement for calibration purposes. The optimum setting was found as ϵeq,SiC = 0.705 ± 0.025 by deduction and as ϵeq,SiC = 0.66 ± 0.06 by experimental validation in the temperature range 120 °C to 400 °C. The apparent temperature coefficient of emissivity, TCE< 2 × 10−4 °C−1 is due to the shift of the Wien peak wavelength relative to the instrument’s sensitivity band.

MetrologyVol. 5(2)
Ford Motor Company (United States) (US), Delft University of Technology (NL)
Openalex Percentile: Top 85%
Thermal Radiation and Cooling Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Setting the Emissivity of an Imaging Bolometer in the Surface Temperature Profile Measurement of SiC-Based MEMS Heaters — David Bilby, R.F. Wolffenbuttel, et al. · Metrology (2025) | TGRS Research Map | TGRS