Structured Light Three-Dimensional Scanning Enables Quantification of Surface Corrosion on Retrieved Spine Rods from Revision Patients

Corrosion of titanium-based metals inside the human body has long been considered a potential health risk. There is an ever-increasing number of spine surgeries performed around the world, and a notable 9% to 45% reported reoperation rate due to spine implant (instrumentation) failure. Instrumentation failure in medicine is not a catastrophic breakage. It is when a patient can no longer biologically or physiologically tolerate the implant; thus, the instrumentation has failed in its function. Most spine reoperations (revisions) are due to a continued progression of spine degeneration or other conditions not associated with infection or breakage. This situation allows the study of how orthopedic implant alloys fare in the human body beyond mechanical failures. Optical inspection is a proven method for area quantification of metallic surface corrosion and wear. Spine instrumentation was retrieved during spine revision procedures under an institution-approved research protocol. Patients who never had any metal implants (n = 6) had an adjacent average Ti tissue concentration of 13.12 μg/g, while patients who had instrumentation (n = 20) had a significantly higher average Ti concentration of 113.73 μg/g. On a per-patient basis (n = 33), the average surface area corrosion was 45.14% while the overall average surface area wear was 6.00%. The presence of a clinically diagnosed infection had a slight but insignificant effect on the overall area of corrosion measured with an increase of corrosion of 38.54% for patients with a clinically diagnosed infection compared to 36.96% for patients with no diagnosed infection. While the 3D scanning with colometric measurements used herein cannot quantify pitting corrosion volume, when combined with gravimetric methods, it may yield a more robust understanding of human in vivo mechanisms of spine rod degradation. The presence of corrosion on all rods in comparison to in vitro studies showing minimal corrosion suggests that our understanding of the human in vivo corrosion environment is still extremely limited.

Authors

Institutions

Publication Details

Journal
CORROSION
Published
2025-06-12
DOI
https://doi.org/10.5006/4708
Primary Topic
Medical Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Structured Light Three-Dimensional Scanning Enables Quantification of Surface Corrosion on Retrieved Spine Rods from Revision Patients

David Ou-Yang, Nolan M. Wessell, Christopher J. Kleck, M. T. Hutchings et al.
CORROSION
Medical Imaging and Analysis
article

Structured Light Three-Dimensional Scanning Enables Quantification of Surface Corrosion on Retrieved Spine Rods from Revision Patients

David Ou-Yang, Nolan M. Wessell, Christopher J. Kleck, M. T. Hutchings, Cheryl L. Ackert‐Bicknell, Michael Rogers, Vikas V. Patel, Reed Ayers, Evalina L. Burger, Julie Albright
article en

Abstract

Corrosion of titanium-based metals inside the human body has long been considered a potential health risk. There is an ever-increasing number of spine surgeries performed around the world, and a notable 9% to 45% reported reoperation rate due to spine implant (instrumentation) failure. Instrumentation failure in medicine is not a catastrophic breakage. It is when a patient can no longer biologically or physiologically tolerate the implant; thus, the instrumentation has failed in its function. Most spine reoperations (revisions) are due to a continued progression of spine degeneration or other conditions not associated with infection or breakage. This situation allows the study of how orthopedic implant alloys fare in the human body beyond mechanical failures. Optical inspection is a proven method for area quantification of metallic surface corrosion and wear. Spine instrumentation was retrieved during spine revision procedures under an institution-approved research protocol. Patients who never had any metal implants (n = 6) had an adjacent average Ti tissue concentration of 13.12 μg/g, while patients who had instrumentation (n = 20) had a significantly higher average Ti concentration of 113.73 μg/g. On a per-patient basis (n = 33), the average surface area corrosion was 45.14% while the overall average surface area wear was 6.00%. The presence of a clinically diagnosed infection had a slight but insignificant effect on the overall area of corrosion measured with an increase of corrosion of 38.54% for patients with a clinically diagnosed infection compared to 36.96% for patients with no diagnosed infection. While the 3D scanning with colometric measurements used herein cannot quantify pitting corrosion volume, when combined with gravimetric methods, it may yield a more robust understanding of human in vivo mechanisms of spine rod degradation. The presence of corrosion on all rods in comparison to in vitro studies showing minimal corrosion suggests that our understanding of the human in vivo corrosion environment is still extremely limited.

CORROSIONVol. 81(7)
University of Denver (US), University of Colorado Anschutz Medical Campus (US), University of Colorado Denver (US)
Openalex Percentile: Top 87%
Medical Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.