Pure high-offset stems can cause an unexpected increase in femoral length in robotic-assisted primary total hip arthroplasty
When pure high-offset stems were used in preoperative planning software, we demonstrated an unexpected increase in leg length between 1-2-mm in 88.4% of patients. This unexpected increase in femoral length is due to a function of the preoperative planning software's planned stem alignment with the anatomical axis, and not an inherent fault in the stem design. With expanding accessibility of robotic-assisted THA platforms, all potential sources of postoperative leg length discrepancy should be identified during preoperative templating and necessary alterations to the surgical plan should be made to accommodate this unexpected difference when using a pure high-offset stem.
Authors
- Aidin Eslam Pour (ORCID: https://orcid.org/0000-0001-7258-1746)
- Claire A. Donnelley (ORCID: https://orcid.org/0000-0001-7316-6368)
- Steven M. Tommasini (ORCID: https://orcid.org/0000-0001-5118-5867)
- Daniel H. Wiznia (ORCID: https://orcid.org/0000-0003-3987-0364)
- Johannes Sieberer (ORCID: https://orcid.org/0000-0002-4191-3442)
- Wei-Shao Tung
Institutions
- Yale University (US)
Publication Details
- Journal
- World Journal of Orthopedics
- Published
- 2025-06-18
- DOI
- https://doi.org/10.5312/wjo.v16.i6.107292
- Primary Topic
- Orthopaedic implants and arthroplasty
- Type
- article
- Field-Weighted Citation Impact
- 0.00