The Influence of the Machining Drill and Direction of Rotation on the Surfaces of Ti6Al4V Dental Implants Subjected to Implantoplasty

Implantoplasty is widely used to treat peri-implantitis by removing biofilms from Ti6Al4V dental implants using rotating drills. This study examined the effects of diamond and tungsten carbide drills, and rotation direction (clockwise/counterclockwise), on surface modification, corrosion behavior, and cytotoxicity. Machining was performed for one minute under a controlled load. Surface roughness, nanohardness, compressive residual stress, and wettability were evaluated, along with SEM and EDX microanalyses of the residues. Corrosion behavior was evaluated using potentiostatic and potentiodynamic tests in Hank's solution. Ion release was monitored over time, and fibroblast viability was tested using extracts at various dilutions. The higher abrasiveness of diamond drills leads to increases roughness from 0.22 mm (control) to 0.73 and 0.59 for diamond and tungsten carbide drills, respectively; in hardness from 2.2 GPa for the control to 4.8 and 3.9 GPa; and in residual compressive stress from -26 to -125 and -111 MPa, with diamond drills inducing more significant changes and producing more hydrophilic surfaces with contact angles around 54° in relation to 80° and 62° for the control and tungsten carbide, respectively. Tungsten carbide drills caused lower corrosion rates (0.0323 mm/year) than diamond drills (0.052 mm/year). In addition, we observed the presence of tungsten ion release. Cytotoxic effects on human fibroblasts were observed with both bur types, and were more pronounced with tungsten carbide, especially at lower dilutions. Only 1:10 dilutions maintained consistent cytocompatibility. The rotation direction showed no significant impact. These findings emphasize the critical influence of bur selection in implantoplasty on the biological response of surrounding tissues.

Authors

Institutions

Publication Details

Journal
Journal of Functional Biomaterials
Published
2025-06-16
DOI
https://doi.org/10.3390/jfb16060224
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Influence of the Machining Drill and Direction of Rotation on the Surfaces of Ti6Al4V Dental Implants Subjected to Implantoplasty

Esteban Padullés-Gaspar, Miquel Punset, Guillermo Cabanes‐Gumbau, Esteban Padullés-Roig et al.
Journal of Functional Biomaterials
Dental Implant Techniques and Outcomes
article

The Influence of the Machining Drill and Direction of Rotation on the Surfaces of Ti6Al4V Dental Implants Subjected to Implantoplasty

Esteban Padullés-Gaspar, Miquel Punset, Guillermo Cabanes‐Gumbau, Esteban Padullés-Roig, Francisco Real‐Voltas, P.B. Fernandez, Javier Gil
article en

Abstract

Implantoplasty is widely used to treat peri-implantitis by removing biofilms from Ti6Al4V dental implants using rotating drills. This study examined the effects of diamond and tungsten carbide drills, and rotation direction (clockwise/counterclockwise), on surface modification, corrosion behavior, and cytotoxicity. Machining was performed for one minute under a controlled load. Surface roughness, nanohardness, compressive residual stress, and wettability were evaluated, along with SEM and EDX microanalyses of the residues. Corrosion behavior was evaluated using potentiostatic and potentiodynamic tests in Hank's solution. Ion release was monitored over time, and fibroblast viability was tested using extracts at various dilutions. The higher abrasiveness of diamond drills leads to increases roughness from 0.22 mm (control) to 0.73 and 0.59 for diamond and tungsten carbide drills, respectively; in hardness from 2.2 GPa for the control to 4.8 and 3.9 GPa; and in residual compressive stress from -26 to -125 and -111 MPa, with diamond drills inducing more significant changes and producing more hydrophilic surfaces with contact angles around 54° in relation to 80° and 62° for the control and tungsten carbide, respectively. Tungsten carbide drills caused lower corrosion rates (0.0323 mm/year) than diamond drills (0.052 mm/year). In addition, we observed the presence of tungsten ion release. Cytotoxic effects on human fibroblasts were observed with both bur types, and were more pronounced with tungsten carbide, especially at lower dilutions. Only 1:10 dilutions maintained consistent cytocompatibility. The rotation direction showed no significant impact. These findings emphasize the critical influence of bur selection in implantoplasty on the biological response of surrounding tissues.

Journal of Functional BiomaterialsVol. 16(6)
Universitat Internacional de Catalunya (ES), Universitat Politècnica de Catalunya (ES)
Openalex Percentile: Top 81%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.