Exploring applications of electrically evoked compound action potentials using the vestibulo-cochlear implant
The presence of a vestibular eCAP was demonstrated to be a predictor of the presence of the eVOR. However, the low negative predictive value prevents the eCAP from providing a reliable indicator for intra-operative electrode position evaluation. eCAP measurements in vestibulo-cochlear configurations indicated that spread of excitation between the cochlea and canals is already possible at stimulation levels within the clinical fitting range of both the vestibular and cochlear electrodes. Measured eCAPs did not correspond with misalignment of the eVOR or the presence of an auditory percept as result of vestibular stimulation. Future research is needed to elucidate the full potential of vestibulo-cochlear implant eCAP measurements.
Authors
- Nils Guinand (ORCID: https://orcid.org/0000-0001-6004-8768)
- Benjamin Volpe (ORCID: https://orcid.org/0009-0007-1125-9295)
- Raymond van de Berg (ORCID: https://orcid.org/0000-0001-9734-8668)
- Bernd L. Vermorken (ORCID: https://orcid.org/0000-0003-2406-8141)
- Stan C J van Boxel (ORCID: https://orcid.org/0000-0002-9647-5056)
- Elke Devocht (ORCID: https://orcid.org/0000-0001-5017-316X)
- Angélica Pérez Fornos (ORCID: https://orcid.org/0000-0003-3911-8826)
Institutions
- Maastricht University Medical Centre (NL)
- Maastricht University (NL)
- University Hospital of Geneva (CH)
Publication Details
- Journal
- Hearing Research
- Published
- 2025-06-16
- DOI
- https://doi.org/10.1016/j.heares.2025.109326
- Citations
- 1
- Primary Topic
- Vestibular and auditory disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.84
Funders
- ZonMw
- Stichting De Weijerhorst