Performance analysis of instrumented socks compared to instrumented treadmill for gait measurements in healthy adults

Traditional gait analysis devices, such as force plates or Instrumented Treadmills (I-TM), often require laboratory settings and can be costly and restrictive, containing their applicability outside controlled environments. To overcome these limitations, we developed the Instrumented Socks Wireless Data Acquisition Module (IS-WDAQ). The accuracy of our developed IS-WDAQ was examined by calculating important temporal parameters (TPs) like step time, stance time, cadence, stance percentage, and swing percentage and compared with I-TM equipment. Participants (n = 250) were divided into groups depending on their age and Body Mass Index (BMI). Categorizing groups by BMI helps to evaluate both systems and ensures accurate, consistent measurements across body types. The Mean Bias Error (MBE) analysis was performed for the two different systems based on their generated TPs. Cadence had the lowest MBE (less than 2 %) among the three conditions tested: Normal (Group N), Overweight (Group OV), and Obese (Group OB). MBE was highest in step time (-5.32 to 7.29 %) and stride time (-9.35 to 4.55 %). For the three BMI groups, stance and swing percentage inaccuracies ranged from -0.78 to 5.18 %. The Intraclass Correlation Coefficient (ICC) showed a very strong cadence association (>0.9) for the three BMI categories across five age groups, while stance and swing percentages were also strongly correlated. The Bland-Altman plot spreads are also within acceptable limits. The study findings imply that the IS-WDAQ is a viable alternative to I-TM for calculating numerous TPs, including cadence, step time, stride time, stance, and swing percentages, especially for use outside laboratory settings.

Authors

Institutions

Publication Details

Journal
Journal of Biomechanics
Published
2025-06-11
DOI
https://doi.org/10.1016/j.jbiomech.2025.112813
Primary Topic
Balance, Gait, and Falls Prevention
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance analysis of instrumented socks compared to instrumented treadmill for gait measurements in healthy adults

T. Sarathi, K. Venkataraman, Amitava Bhattacharyya, B Dinakar
Journal of Biomechanics
Balance, Gait, and Falls Prevention
article

Performance analysis of instrumented socks compared to instrumented treadmill for gait measurements in healthy adults

T. Sarathi, K. Venkataraman, Amitava Bhattacharyya, B Dinakar
article en

Abstract

Traditional gait analysis devices, such as force plates or Instrumented Treadmills (I-TM), often require laboratory settings and can be costly and restrictive, containing their applicability outside controlled environments. To overcome these limitations, we developed the Instrumented Socks Wireless Data Acquisition Module (IS-WDAQ). The accuracy of our developed IS-WDAQ was examined by calculating important temporal parameters (TPs) like step time, stance time, cadence, stance percentage, and swing percentage and compared with I-TM equipment. Participants (n = 250) were divided into groups depending on their age and Body Mass Index (BMI). Categorizing groups by BMI helps to evaluate both systems and ensures accurate, consistent measurements across body types. The Mean Bias Error (MBE) analysis was performed for the two different systems based on their generated TPs. Cadence had the lowest MBE (less than 2 %) among the three conditions tested: Normal (Group N), Overweight (Group OV), and Obese (Group OB). MBE was highest in step time (-5.32 to 7.29 %) and stride time (-9.35 to 4.55 %). For the three BMI groups, stance and swing percentage inaccuracies ranged from -0.78 to 5.18 %. The Intraclass Correlation Coefficient (ICC) showed a very strong cadence association (>0.9) for the three BMI categories across five age groups, while stance and swing percentages were also strongly correlated. The Bland-Altman plot spreads are also within acceptable limits. The study findings imply that the IS-WDAQ is a viable alternative to I-TM for calculating numerous TPs, including cadence, step time, stride time, stance, and swing percentages, especially for use outside laboratory settings.

Journal of BiomechanicsVol. 189
Seoul National University of Science and Technology (KR), KG Hospital (IN), PSG Institute of Advanced Studies (IN)
Department of Science and Technology, Ministry of Science and Technology, India
Zero hunger
Openalex Percentile: Top 78%
Balance, Gait, and Falls Prevention
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.