Comparison of the Optical Biometry Based on Swept‐Source Optical Coherence Tomography and Ultrasound‐Based Biometry Device in Patients With Mature Cataract

Objectives: To compare the agreement and reliability values between an optical biometry based on swept‐source optical coherence tomography (OCT) (IOLMaster 700; Carl Zeiss, Germany) and an ocular ultrasound‐based biometry device (Nidek; US‐4000 Echoscan, Japan) for axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements. Methods: In this cross‐sectional study, 101 eyes of 101 consecutive patients with mature cataract underwent routine eye examinations. Then, ocular biometry measurements through optical biometry and contact ultrasound biometry were carried out. The intraclass correlation coefficient (ICC) for reliability and Bland–Altman plots for the agreement between the mean findings of the two devices were analyzed. Results: The mean age of patients was 61.6 ± 13.1 (range, 27–94) years. For AL, the mean differences were small (0.10 mm), but with a very strong correlation and excellent agreement ( r = 0.999, ICC = 0.999, p < 0.001) between the two devices. Similarly, for ACD, the correlation was very high with excellent reliability ( r = 0.958, ICC = 0.974, p < 0.001), though the ultrasound biometer reported a slightly deeper mean ACD (by 0.13 mm). For LT, the agreement and correlation were lower ( r = 0.508, ICC = 0.663, p < 0.001), with the optical biometer reporting a thicker mean LT (0.31 mm) than the ultrasound biometer. Conclusion: AL and ACD measurements from the optical biometry based on swept‐source OCT and ultrasound can be used interchangeably in patients with mature cataract. However, LT measurements from the two devices should not be considered interchangeable.

Authors

Institutions

Publication Details

Journal
International Journal of Clinical Practice
Published
2025-01-01
DOI
https://doi.org/10.1155/ijcp/5679146
Primary Topic
Retinal Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of the Optical Biometry Based on Swept‐Source Optical Coherence Tomography and Ultrasound‐Based Biometry Device in Patients With Mature Cataract

Mahsa Ranjbar-Pazooki, Elham Azizi, Gaber Ali, Ali Majdi et al.
International Journal of Clinical Practice
Retinal Imaging and Analysis
article

Comparison of the Optical Biometry Based on Swept‐Source Optical Coherence Tomography and Ultrasound‐Based Biometry Device in Patients With Mature Cataract

Mahsa Ranjbar-Pazooki, Elham Azizi, Gaber Ali, Ali Majdi, Masoud Khorrami‐Nejad, Mehdi Khodaparast
article en

Abstract

Objectives: To compare the agreement and reliability values between an optical biometry based on swept‐source optical coherence tomography (OCT) (IOLMaster 700; Carl Zeiss, Germany) and an ocular ultrasound‐based biometry device (Nidek; US‐4000 Echoscan, Japan) for axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements. Methods: In this cross‐sectional study, 101 eyes of 101 consecutive patients with mature cataract underwent routine eye examinations. Then, ocular biometry measurements through optical biometry and contact ultrasound biometry were carried out. The intraclass correlation coefficient (ICC) for reliability and Bland–Altman plots for the agreement between the mean findings of the two devices were analyzed. Results: The mean age of patients was 61.6 ± 13.1 (range, 27–94) years. For AL, the mean differences were small (0.10 mm), but with a very strong correlation and excellent agreement ( r = 0.999, ICC = 0.999, p < 0.001) between the two devices. Similarly, for ACD, the correlation was very high with excellent reliability ( r = 0.958, ICC = 0.974, p < 0.001), though the ultrasound biometer reported a slightly deeper mean ACD (by 0.13 mm). For LT, the agreement and correlation were lower ( r = 0.508, ICC = 0.663, p < 0.001), with the optical biometer reporting a thicker mean LT (0.31 mm) than the ultrasound biometer. Conclusion: AL and ACD measurements from the optical biometry based on swept‐source OCT and ultrasound can be used interchangeably in patients with mature cataract. However, LT measurements from the two devices should not be considered interchangeable.

International Journal of Clinical PracticeVol. 2025(1)
Alsalam University College (IQ), Farabi Hospital (IR), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 79%
Retinal Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.