Choosing the Correct Internal Reference Redox Species for Overcoming Reference Electrode Drift in Voltammetric pH Measurements

Reference electrode (RE) drift is a common problem when electrodes are used for pH determination, especially over extended periods of time or in complex media. For voltammetric pH measurements, one method to mitigate against RE drift is to add a second pH insensitive redox species (the internal reference, IREF) and measure the difference in peak potential, <i>E</i> <sub>diff</sub>, between the signal associated with the pH sensitive species, <i>E</i> <sub>pH</sub>, and IREF, <i>E</i> <sub>IREF</sub>. This work strategically explores how to choose the correct IREF species. For these studies, a quinone-functionalized boron doped diamond (BDD-Q) electrode is employed as the pH sensing electrode over the pH range of 4-9. To avoid errors in reporting of the real <i>E</i> <sub>pH</sub> and <i>E</i> <sub>IREF</sub> values, there must be a minimum separation between the two peaks. Moreover, the distance on the potential axis between a peak and that of the current response pertaining to water electrolysis must also be considered. For the BDD-Q pH electrode, an operable potential window for IREF is established and the IREF redox species, hexachloroiridate (IrCl<sub>6</sub> <sup>2-/3-</sup>), is determined to be most appropriate, showing an ∼0.08 pH error over the pH range of 4-9 and reducing to ∼0.02 pH error over the pH range of 6-8. The use of <i>E</i> <sub>diff</sub> is further assessed via the voltammetric measurement of dissolved carbon dioxide (CO<sub>2</sub>) in a Stow-Severinghaus arrangement over the partial pressure range of 30.4-152.0 mmHg. The <i>R</i> <sup>2</sup> linearity of the calibration line (=0.998) is shown to be equivalent and in agreement with theory when plotting either <i>E</i> <sub>diff</sub> or <i>E</i> <sub>pH</sub> versus CO<sub>2</sub> partial pressure. This data bodes well for the use of <i>E</i> <sub>diff</sub> as a measurement signal in Stow-Severinghaus dissolved CO<sub>2</sub> transcutaneous sensors, where continuous measurement of pH over several days is required.

Authors

Institutions

Publication Details

Journal
ACS electrochemistry.
Published
2025-06-13
DOI
https://doi.org/10.1021/acselectrochem.5c00138
Citations
3
Primary Topic
Electrochemical Analysis and Applications
Type
article
Field-Weighted Citation Impact
1.55

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Choosing the Correct Internal Reference Redox Species for Overcoming Reference Electrode Drift in Voltammetric pH Measurements

Katherine J. Levey, Tania L. Read, Julie V. Macpherson, N. Biswas
3 citations
ACS electrochemistry.
Electrochemical Analysis and Applications
1.55
article

Choosing the Correct Internal Reference Redox Species for Overcoming Reference Electrode Drift in Voltammetric pH Measurements

Katherine J. Levey, Tania L. Read, Julie V. Macpherson, N. Biswas
article en
3 citations

Abstract

Reference electrode (RE) drift is a common problem when electrodes are used for pH determination, especially over extended periods of time or in complex media. For voltammetric pH measurements, one method to mitigate against RE drift is to add a second pH insensitive redox species (the internal reference, IREF) and measure the difference in peak potential, <i>E</i> <sub>diff</sub>, between the signal associated with the pH sensitive species, <i>E</i> <sub>pH</sub>, and IREF, <i>E</i> <sub>IREF</sub>. This work strategically explores how to choose the correct IREF species. For these studies, a quinone-functionalized boron doped diamond (BDD-Q) electrode is employed as the pH sensing electrode over the pH range of 4-9. To avoid errors in reporting of the real <i>E</i> <sub>pH</sub> and <i>E</i> <sub>IREF</sub> values, there must be a minimum separation between the two peaks. Moreover, the distance on the potential axis between a peak and that of the current response pertaining to water electrolysis must also be considered. For the BDD-Q pH electrode, an operable potential window for IREF is established and the IREF redox species, hexachloroiridate (IrCl<sub>6</sub> <sup>2-/3-</sup>), is determined to be most appropriate, showing an ∼0.08 pH error over the pH range of 4-9 and reducing to ∼0.02 pH error over the pH range of 6-8. The use of <i>E</i> <sub>diff</sub> is further assessed via the voltammetric measurement of dissolved carbon dioxide (CO<sub>2</sub>) in a Stow-Severinghaus arrangement over the partial pressure range of 30.4-152.0 mmHg. The <i>R</i> <sup>2</sup> linearity of the calibration line (=0.998) is shown to be equivalent and in agreement with theory when plotting either <i>E</i> <sub>diff</sub> or <i>E</i> <sub>pH</sub> versus CO<sub>2</sub> partial pressure. This data bodes well for the use of <i>E</i> <sub>diff</sub> as a measurement signal in Stow-Severinghaus dissolved CO<sub>2</sub> transcutaneous sensors, where continuous measurement of pH over several days is required.

ACS electrochemistry.Vol. 1(8)
Leiden University (NL), University of Warwick (GB)
Medical Research Council, Engineering and Physical Sciences Research Council
Openalex Percentile: Top 19%
Electrochemical Analysis and Applications
1.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.