Preliminary evaluation of smartphone-based addition measurement in a presbyopic population
Linear regression analysis showed a strong correlation (R² = 0.82) between app-measured and clinical addition values. Bland-Altman analysis revealed a mean over estimation of -0.22± 0.38 D with a limit of agreement of ±0.74 D of the near correction. Cumulative error analysis indicated that 61.7 % of app measurements were within 0.25 D of clinical values, and 82.5 % within 0.50 D CONCLUSIONS: A smartphone-based subjective measurement of presbyopic addition can potentially be used to assess the addition needed to detect the under or over corrected addition in full presbyopic subjects.
Authors
- Norberto López‐Gil (ORCID: https://orcid.org/0000-0002-6696-4596)
- Rosa María Salmerón-Campillo (ORCID: https://orcid.org/0000-0001-5710-8555)
- Matt Jaskulski (ORCID: https://orcid.org/0000-0001-5824-7135)
- Ginés Martínez-Ros
- José Ángel Díaz-Guirado
Institutions
- Indiana University Bloomington (US)
- Universidad de Murcia (ES)
Publication Details
- Journal
- Journal of Optometry
- Published
- 2025-06-13
- DOI
- https://doi.org/10.1016/j.optom.2025.100561
- Citations
- 2
- Primary Topic
- Gaze Tracking and Assistive Technology
- Type
- article
- Field-Weighted Citation Impact
- 3.87