Preprocedural 3D Transesophageal Echocardiography for the Prediction of Device Deformation Morphology and Peri-Device Leaks After Transcatheter Left Atrial Appendage Occlusion with the AmplatzerTM Device

<b>Background/Objectives</b>: Percutaneous left atrial appendage occlusion (LAAO) has become an alternative to oral anticoagulation in selected patients with atrial fibrillation. The results of transcatheter LAAO were evaluated by conventional cardiac angiography (CCA), fluoroscopy, and 2D and 3D transesophageal echocardiography (TEE). <b>Methods</b>: In 47 consecutive patients (76 ± 8 years), LAAO was performed with the Amplatzer<sup>TM</sup> device. CCA and 2D and 3D TEE were performed for LAA measurements. The eccentricity of the device landing zone was assessed by 3D TEE as the ratio of the maximal and minimal diameters. The device size was selected depending on the CCA maximal diameter. The postprocedural device lobe morphology was classified by fluoroscopy as "square" type (sign of undersizing), "tire" type (optimal deformation), and "strawberry" type (sign of oversizing). After 6 months, TEE was repeated to evaluate peri-device leaks (PDLs). <b>Results</b>: The postprocedural device morphology was "square" type in 9 (19%), "tire" type in 28 (60%), and "strawberry" type in 10 patients (21%). After 6 months, peri-device leaks were observed in 11 patients (23%), with the lowest incidence in the "tire"-type group (11%) compared with "square" type (56%) and "strawberry" type (30%) (<i>p</i> = 0.019). The 3D TEE eccentricity index with a cut-off value of ≤1.29 (indicating a more circular morphology) was a very specific predictor for excluding postprocedural device deformation of the "strawberry" type (AUC = 0.689; sensitivity 48.7%, specificity 100%). <b>Conclusions</b>: Undersizing as well as oversizing of the LAA occluder, as characterized by deformation type by fluoroscopy, was associated with postprocedural PDLs. The eccentricity of the LAA landing zone by 3D TEE may indicate inadequate size selection of the Amplatzer<sup>TM</sup> device leading to oversizing in patients with a more eccentric LAA.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-06-13
DOI
https://doi.org/10.3390/jcm14124211
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Preprocedural 3D Transesophageal Echocardiography for the Prediction of Device Deformation Morphology and Peri-Device Leaks After Transcatheter Left Atrial Appendage Occlusion with the AmplatzerTM Device

Ertunc Altiok, Nikolaus Marx, Felix Vogt, Michael Frick et al.
Journal of Clinical Medicine
Atrial Fibrillation Management and Outcomes
article

Preprocedural 3D Transesophageal Echocardiography for the Prediction of Device Deformation Morphology and Peri-Device Leaks After Transcatheter Left Atrial Appendage Occlusion with the AmplatzerTM Device

Ertunc Altiok, Nikolaus Marx, Felix Vogt, Michael Frick, Jörg Schroeder, Rami Al-khusein, Andreas Pütz, Annemarie Kirschfink, Kinan Kneizeh, Anas Alnaimi, Pieterjan van Rijckeghem
article en

Abstract

<b>Background/Objectives</b>: Percutaneous left atrial appendage occlusion (LAAO) has become an alternative to oral anticoagulation in selected patients with atrial fibrillation. The results of transcatheter LAAO were evaluated by conventional cardiac angiography (CCA), fluoroscopy, and 2D and 3D transesophageal echocardiography (TEE). <b>Methods</b>: In 47 consecutive patients (76 ± 8 years), LAAO was performed with the Amplatzer<sup>TM</sup> device. CCA and 2D and 3D TEE were performed for LAA measurements. The eccentricity of the device landing zone was assessed by 3D TEE as the ratio of the maximal and minimal diameters. The device size was selected depending on the CCA maximal diameter. The postprocedural device lobe morphology was classified by fluoroscopy as "square" type (sign of undersizing), "tire" type (optimal deformation), and "strawberry" type (sign of oversizing). After 6 months, TEE was repeated to evaluate peri-device leaks (PDLs). <b>Results</b>: The postprocedural device morphology was "square" type in 9 (19%), "tire" type in 28 (60%), and "strawberry" type in 10 patients (21%). After 6 months, peri-device leaks were observed in 11 patients (23%), with the lowest incidence in the "tire"-type group (11%) compared with "square" type (56%) and "strawberry" type (30%) (<i>p</i> = 0.019). The 3D TEE eccentricity index with a cut-off value of ≤1.29 (indicating a more circular morphology) was a very specific predictor for excluding postprocedural device deformation of the "strawberry" type (AUC = 0.689; sensitivity 48.7%, specificity 100%). <b>Conclusions</b>: Undersizing as well as oversizing of the LAA occluder, as characterized by deformation type by fluoroscopy, was associated with postprocedural PDLs. The eccentricity of the LAA landing zone by 3D TEE may indicate inadequate size selection of the Amplatzer<sup>TM</sup> device leading to oversizing in patients with a more eccentric LAA.

Journal of Clinical MedicineVol. 14(12)
RWTH Aachen University (DE)
Ernst und Berta Grimmke Stiftung
Openalex Percentile: Top 76%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.