Comparing respiratory variables with circle, Bain and T-piece breathing systems in 5–10 kg dogs: a randomized crossover trial
While differences were found between both test systems and the control, the clinical consequences appear minimal. Both circle and Bain, appear suitable for use in healthy 5-10 kg dogs as overall ventilatory function was adequate.