Biomechanical Effects of Partial Decortication on All-Suture and Conventional Suture Anchors in Different Bone Densities

While biomechanical testing has shown a correlation between decortication and anchor failure load, the effects of partial decortication on the biomechanical properties of all-suture anchors remain unclear. We aimed to evaluate the biomechanical effects of partial decortication on all-suture anchors and conventional suture anchors in Sawbones of varying densities. Suture anchors were tested in nondecorticated, partially decorticated, and completely decorticated Sawbones. Two types of all-suture anchors and one type of conventional anchor were evaluated. Two types of biphasic polyurethane foam were used to mimic normal bone: 0.32 g/cm<sup>3</sup> density (20 pounds per cubic foot, pcf 20) and osteoporotic bone: 0.16 g/cm<sup>3</sup> density (10 pounds per cubic foot, pcf 10). Cyclic loadings were applied, and peak displacement was recorded. After cyclic loading tests, surviving anchors were subjected to pull-to-failure tests. The number of cycles, peak displacement, ultimate failure loads, and failure modes were determined. First, peak displacement was significantly influenced by bone density and anchor type: normal bone models exhibited lower peak displacement than osteoporotic models, and conventional screw-type anchors consistently demonstrated reduced peak displacement compared to all-suture anchors. In contrast, the extent of bone decortication-whether non-decorticated, partially decorticated, or completely decorticated-showed no significant effect on peak displacement. Second, in osteoporotic bone models (10 pounds per cubic foot), no significant difference in failure load was observed between the partially and non-decorticated groups, but both exhibited significantly higher values than the completely decorticated group.

Authors

Institutions

Publication Details

Journal
Journal of Visualized Experiments
Published
2025-06-13
DOI
https://doi.org/10.3791/68355
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical Effects of Partial Decortication on All-Suture and Conventional Suture Anchors in Different Bone Densities

Yang Zhao, Chi Xiao, Chunyan Jiang, Yi-ming Zhu et al.
Journal of Visualized Experiments
Shoulder Injury and Treatment
article

Biomechanical Effects of Partial Decortication on All-Suture and Conventional Suture Anchors in Different Bone Densities

Yang Zhao, Chi Xiao, Chunyan Jiang, Yi-ming Zhu, Xiao Lian
article en

Abstract

While biomechanical testing has shown a correlation between decortication and anchor failure load, the effects of partial decortication on the biomechanical properties of all-suture anchors remain unclear. We aimed to evaluate the biomechanical effects of partial decortication on all-suture anchors and conventional suture anchors in Sawbones of varying densities. Suture anchors were tested in nondecorticated, partially decorticated, and completely decorticated Sawbones. Two types of all-suture anchors and one type of conventional anchor were evaluated. Two types of biphasic polyurethane foam were used to mimic normal bone: 0.32 g/cm<sup>3</sup> density (20 pounds per cubic foot, pcf 20) and osteoporotic bone: 0.16 g/cm<sup>3</sup> density (10 pounds per cubic foot, pcf 10). Cyclic loadings were applied, and peak displacement was recorded. After cyclic loading tests, surviving anchors were subjected to pull-to-failure tests. The number of cycles, peak displacement, ultimate failure loads, and failure modes were determined. First, peak displacement was significantly influenced by bone density and anchor type: normal bone models exhibited lower peak displacement than osteoporotic models, and conventional screw-type anchors consistently demonstrated reduced peak displacement compared to all-suture anchors. In contrast, the extent of bone decortication-whether non-decorticated, partially decorticated, or completely decorticated-showed no significant effect on peak displacement. Second, in osteoporotic bone models (10 pounds per cubic foot), no significant difference in failure load was observed between the partially and non-decorticated groups, but both exhibited significantly higher values than the completely decorticated group.

Journal of Visualized Experiments(220)
Chinese Academy of Sciences (CN), Peking University (CN), Beijing Jishuitan Hospital (CN), Institute of Mechanics (CN)
Zero hunger
Openalex Percentile: Top 77%
Shoulder Injury and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.