Radiation protection in practice: assessment of shielding block efficacy in external beam radiotherapy at a leading cancer centre in sub-Saharan Africa

Abstract Background: Radiotherapy aims to provide either curative treatment or palliation, with radiation transmission through shielding blocks ideally not exceeding 5% of the incident dose. Aim: This study assessed the attenuation efficacy of Cerrobend shielding blocks, determining the thickness required for achieving 95% attenuation and evaluating the adequacy of currently used blocks. Methods: An experimental study was conducted using a 1·25 MeV cobalt-60 teletherapy machine. Radiation intensities transmitted through Cerrobend blocks of varying thicknesses (1–5 cm) were measured with a calibrated ionization chamber and electrometer. Measurements were performed with a fixed field size of 10 cm × 10 cm at a 100-cm source-to-chamber distance, minimizing scatter. Baseline and transmitted intensities were recorded, with each measurement repeated five times for precision. Results: The percentage transmission, linear attenuation coefficients and thickness required for 5% transmission were calculated using regression analysis. Radiation intensity decreased exponentially with increasing Cerrobend block thickness, from 15·01 nC (0 cm) to 0·939 nC (5 cm thickness). Regression analysis showed a strong negative linear relationship (y = −0·5386x + 2·6705; R 2 = 0·992). The thickness required for maximum transmission of 5% was determined to be 6 cm, while the routinely used 5 cm blocks at the study site allowed 8% transmission, exceeding the recommended threshold. Minimal environmental variations ensured measurement consistency. Conclusion: The study demonstrated that 6 cm Cerrobend blocks were necessary to achieve the recommended 95% attenuation, whereas the 5 cm blocks in use resulted in suboptimal shielding. This highlights the need to revise shielding practices to improve radiation protection and patient safety in resource-constrained radiotherapy settings.

Authors

Institutions

Publication Details

Journal
Journal of Radiotherapy in Practice
Published
2025-01-01
DOI
https://doi.org/10.1017/s1460396925100125
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation protection in practice: assessment of shielding block efficacy in external beam radiotherapy at a leading cancer centre in sub-Saharan Africa

Kofi Adesi Kyei, Joseph P. Daniels
Journal of Radiotherapy in Practice
Advanced Radiotherapy Techniques
article

Radiation protection in practice: assessment of shielding block efficacy in external beam radiotherapy at a leading cancer centre in sub-Saharan Africa

Kofi Adesi Kyei, Joseph P. Daniels
article en

Abstract

Abstract Background: Radiotherapy aims to provide either curative treatment or palliation, with radiation transmission through shielding blocks ideally not exceeding 5% of the incident dose. Aim: This study assessed the attenuation efficacy of Cerrobend shielding blocks, determining the thickness required for achieving 95% attenuation and evaluating the adequacy of currently used blocks. Methods: An experimental study was conducted using a 1·25 MeV cobalt-60 teletherapy machine. Radiation intensities transmitted through Cerrobend blocks of varying thicknesses (1–5 cm) were measured with a calibrated ionization chamber and electrometer. Measurements were performed with a fixed field size of 10 cm × 10 cm at a 100-cm source-to-chamber distance, minimizing scatter. Baseline and transmitted intensities were recorded, with each measurement repeated five times for precision. Results: The percentage transmission, linear attenuation coefficients and thickness required for 5% transmission were calculated using regression analysis. Radiation intensity decreased exponentially with increasing Cerrobend block thickness, from 15·01 nC (0 cm) to 0·939 nC (5 cm thickness). Regression analysis showed a strong negative linear relationship (y = −0·5386x + 2·6705; R 2 = 0·992). The thickness required for maximum transmission of 5% was determined to be 6 cm, while the routinely used 5 cm blocks at the study site allowed 8% transmission, exceeding the recommended threshold. Minimal environmental variations ensured measurement consistency. Conclusion: The study demonstrated that 6 cm Cerrobend blocks were necessary to achieve the recommended 95% attenuation, whereas the 5 cm blocks in use resulted in suboptimal shielding. This highlights the need to revise shielding practices to improve radiation protection and patient safety in resource-constrained radiotherapy settings.

Journal of Radiotherapy in PracticeVol. 24
University of Ghana (GH), Korle Bu Teaching Hospital (GH)
Openalex Percentile: Top 81%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.