Interfering heterophile antibodies as the cause of persistently falsely elevated high-sensitivity troponin I on Alinity i: a case report

Hereby we describe a case of a 59-year-old female patient with persistently elevated high-sensitivity troponin I (hs-TnI) over the course of almost four years measured on Alinity i with the corresponding assay (Abbott Laboratories, Chicago, USA). The patient underwent multiple extensive cardiological evaluations, but none of them suggested acute or chronic cardiac damage. Therefore, interference in measurement was suspected by the attending cardiologist and a detailed, stepwise laboratory investigation was undertaken in the sample with initial hs-TnI result of 2077 ng/L. Serial sample dilutions (1:2,1:5,1:10) did not match the expected, calculated hs-TnI concentrations, yielding both huge positive biases (62, 109 and 139%, respectively) and absolute differences (639, 453 and 290 ng/L, respectively). Precipitation with polyethylene-glycol, pretreatment in heterophilic blocking tubes (HBT) and immunoglobulin G depletion yielded hs-TnI results below the assay's diagnostic cut-off (< 15.6 ng/L). Alternate hs-TnI immunoassays (Siemens Healthineers, Beckman Coulter and Snibe) and measurement with the high-sensitivity troponin T (hs-TnT) assay yielded results below assays' specific cut-off values. This investigation confirmed that results of hs-TnI obtained by the Abbott assay were spuriously elevated. Significant lowering of hs-TnI after HBT pretreatment indicated that heterophile antibodies are the most probable source of interference. Based on this finding, it was entered in the patient's medical record that future determinations of cardiac troponin should be performed with an alternate hs-TnI or hs-TnT assay. This case emphasizes that analytical interferences are usually immunoassay-dependent. Evaluation of laboratory results in the clinical context and close collaboration between laboratory and clinical staff is crucial for their recognition.

Authors

Institutions

Publication Details

Journal
Biochemia Medica
Published
2025-06-11
DOI
https://doi.org/10.11613/bm.2025.021001
Primary Topic
Acute Myocardial Infarction Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interfering heterophile antibodies as the cause of persistently falsely elevated high-sensitivity troponin I on Alinity i: a case report

Sandra Jakšić Jurinjak, Dragana Šegulja, Lovorka Đerek, Željka Vogrinc et al.
Biochemia Medica
Acute Myocardial Infarction Research
article

Interfering heterophile antibodies as the cause of persistently falsely elevated high-sensitivity troponin I on Alinity i: a case report

Sandra Jakšić Jurinjak, Dragana Šegulja, Lovorka Đerek, Željka Vogrinc, Sanja Kačkov, Fran Smaić, Ivana Lapić, Dunja Rogić, Indira Imširović
article en

Abstract

Hereby we describe a case of a 59-year-old female patient with persistently elevated high-sensitivity troponin I (hs-TnI) over the course of almost four years measured on Alinity i with the corresponding assay (Abbott Laboratories, Chicago, USA). The patient underwent multiple extensive cardiological evaluations, but none of them suggested acute or chronic cardiac damage. Therefore, interference in measurement was suspected by the attending cardiologist and a detailed, stepwise laboratory investigation was undertaken in the sample with initial hs-TnI result of 2077 ng/L. Serial sample dilutions (1:2,1:5,1:10) did not match the expected, calculated hs-TnI concentrations, yielding both huge positive biases (62, 109 and 139%, respectively) and absolute differences (639, 453 and 290 ng/L, respectively). Precipitation with polyethylene-glycol, pretreatment in heterophilic blocking tubes (HBT) and immunoglobulin G depletion yielded hs-TnI results below the assay's diagnostic cut-off (< 15.6 ng/L). Alternate hs-TnI immunoassays (Siemens Healthineers, Beckman Coulter and Snibe) and measurement with the high-sensitivity troponin T (hs-TnT) assay yielded results below assays' specific cut-off values. This investigation confirmed that results of hs-TnI obtained by the Abbott assay were spuriously elevated. Significant lowering of hs-TnI after HBT pretreatment indicated that heterophile antibodies are the most probable source of interference. Based on this finding, it was entered in the patient's medical record that future determinations of cardiac troponin should be performed with an alternate hs-TnI or hs-TnT assay. This case emphasizes that analytical interferences are usually immunoassay-dependent. Evaluation of laboratory results in the clinical context and close collaboration between laboratory and clinical staff is crucial for their recognition.

Biochemia MedicaVol. 35(2)
University Hospital Centre Zagreb (HR), University of Zagreb (HR), University Hospital Dubrava (HR)
Good health and well-being
Openalex Percentile: Top 76%
Acute Myocardial Infarction Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.