Application of the Brillouin Optical Scanning System in the Regional Corneal Biomechanical Evaluation of Keratoconus and Its Correlation with Corvis ST Parameters

(1) Background: The early diagnosis of keratoconus is critical for prognosis. Traditional methods like ORA and Corvis ST measure overall corneal biomechanics but lack regional specificity and are affected by intraocular pressure. In contrast, Brillouin microscopy assesses regional corneal biomechanics without such limitations; (2) Methods: In total, 25 keratoconus patients and 28 healthy controls were included in this study. Corneal biomechanics were measured using the BOSS system (Brillouin Optical Scanning System) in a 10-point mode within an 8 mm diameter, and included the mean, maximum, minimum and standard Brillouin shift. The Corvis ST parameters extracted included the CBI (Corneal Biomechanical Index), CCBI (Corvis Biomechanical Index for Chinese populations), SSI (Stress-Strain Index), DA (Deformation Amplitude), IIR (Inverse Integrated Radius), and SP-A1 (Stiffness Parameter at First Applanation); (3) Results: BOSS showed significant differences in the inferior nasal region (<i>p</i> = 0.004) and central region (<i>p</i> = 0.029) between groups, but not in peripheral regions (<i>p</i> = 0.781). In a comparison of the Brillouin frequency shifts measured between groups, there was no difference in the Mean (<i>p</i> = 0.452) and Max (<i>p</i> = 0.487), but the Min (<i>p</i> = 0.003), Standard (<i>p</i> = 0.000), and Max-Min (<i>p</i> = 0.006) all showed differences. Corvis ST identified significant differences in six parameters (CBI, CCBI, SSI, DA, IIR, and SP-A1) between groups (<i>p</i> < 0.001). Correlations were found between the BOSS and Corvis ST results, with moderate correlations in the inferior nasal region; (4) Conclusions: The BOSS Brillouin microscope can provide an accurate diagnostic evaluation for the corneal biomechanical differences between normal eyes and keratoconus, independent of IOP (Intraocular Pressure) and CCT (Central Corneal Thickness), with a good correlation with Corvis ST, especially in assessing regional biomechanics.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-06-11
DOI
https://doi.org/10.3390/bioengineering12060634
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Application of the Brillouin Optical Scanning System in the Regional Corneal Biomechanical Evaluation of Keratoconus and Its Correlation with Corvis ST Parameters

Jialu Chen, Lumeng Wang, Shihao Chen, Zhiguang Gu et al.
Bioengineering
Corneal surgery and disorders
article

Application of the Brillouin Optical Scanning System in the Regional Corneal Biomechanical Evaluation of Keratoconus and Its Correlation with Corvis ST Parameters

Jialu Chen, Lumeng Wang, Shihao Chen, Zhiguang Gu, Yiqiang Wu, Guangkai Sun, Zhiyi Chen, Qiuruo Jiang, XiaoBo Zheng
article en

Abstract

(1) Background: The early diagnosis of keratoconus is critical for prognosis. Traditional methods like ORA and Corvis ST measure overall corneal biomechanics but lack regional specificity and are affected by intraocular pressure. In contrast, Brillouin microscopy assesses regional corneal biomechanics without such limitations; (2) Methods: In total, 25 keratoconus patients and 28 healthy controls were included in this study. Corneal biomechanics were measured using the BOSS system (Brillouin Optical Scanning System) in a 10-point mode within an 8 mm diameter, and included the mean, maximum, minimum and standard Brillouin shift. The Corvis ST parameters extracted included the CBI (Corneal Biomechanical Index), CCBI (Corvis Biomechanical Index for Chinese populations), SSI (Stress-Strain Index), DA (Deformation Amplitude), IIR (Inverse Integrated Radius), and SP-A1 (Stiffness Parameter at First Applanation); (3) Results: BOSS showed significant differences in the inferior nasal region (<i>p</i> = 0.004) and central region (<i>p</i> = 0.029) between groups, but not in peripheral regions (<i>p</i> = 0.781). In a comparison of the Brillouin frequency shifts measured between groups, there was no difference in the Mean (<i>p</i> = 0.452) and Max (<i>p</i> = 0.487), but the Min (<i>p</i> = 0.003), Standard (<i>p</i> = 0.000), and Max-Min (<i>p</i> = 0.006) all showed differences. Corvis ST identified significant differences in six parameters (CBI, CCBI, SSI, DA, IIR, and SP-A1) between groups (<i>p</i> < 0.001). Correlations were found between the BOSS and Corvis ST results, with moderate correlations in the inferior nasal region; (4) Conclusions: The BOSS Brillouin microscope can provide an accurate diagnostic evaluation for the corneal biomechanical differences between normal eyes and keratoconus, independent of IOP (Intraocular Pressure) and CCT (Central Corneal Thickness), with a good correlation with Corvis ST, especially in assessing regional biomechanics.

BioengineeringVol. 12(6)
Wenzhou Medical University (CN), Affiliated Eye Hospital of Wenzhou Medical College (CN), University College London (GB)
Openalex Percentile: Top 80%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.