Supplemental cerclage wiring at the distal tibia has shown to improve the mechanical stabilization of the fracture. However, its clinical use remains controversial partly because of concerns about implant loosening as well as frictional or constrictive interference between the cerclage and cortical surface. The aim of this study was to investigate in a distal tibia fracture model possible loosening of the cerclage and interference with the bone surface. Fourteen distal tibia oblique fractures (AO/OTA 42-A2) in seven human and synthetic bone samples, respectively were treated by plate osteosynthesis with supplemental cerclage wiring. The samples were subjected to 50,000 load cycles under combined axial and torsional loads. Angular and axial displacement were continuously monitored to identify possible cerclage loosening cerclage. Potential incision of the cerclage was inspected visually and on CT scans. Angular displacement was significantly influenced by the bone material (p ≤ 0.001) and type of osteosynthesis (p ≤ 0.001), while axial displacement was only influenced by the type of osteosynthesis (p ≤ 0.001). Lowest displacements were found in samples with plate and cable cerclage. Cerclage stability was maintained throughout the entire test period of 50,000 load cycles for human and synthetic samples. CT images revealed notches from the cerclage clamping mechanism but no incisions of the cable itself. In an oblique distal tibia fracture model, loosening of the cable cerclage was not detected. Full weight-bearing cyclic loads resulted in slight imprints of the cerclage crimp in both, human and synthetic samples. Following the surgical guidelines for careful cerclage installation, a supplemental cable cerclage has the potential to significantly increase the construct stability.

Authors

Institutions

Publication Details

Journal
European Journal of Trauma and Emergency Surgery
Published
2025-06-11
DOI
https://doi.org/10.1007/s00068-025-02894-8
Primary Topic
Bone fractures and treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical evaluation of cerclage wiring in plated tibia fractures using human and synthetic specimens

Marianne Hollensteiner, Edgar Mayr, Peter Augat, Sabrina Sandriesser et al.
European Journal of Trauma and Emergency Surgery
Bone fractures and treatments
article

Biomechanical evaluation of cerclage wiring in plated tibia fractures using human and synthetic specimens

Marianne Hollensteiner, Edgar Mayr, Peter Augat, Sabrina Sandriesser, Stefan Förch, Jan Reuter, Christoph Kern
article en

Abstract

Supplemental cerclage wiring at the distal tibia has shown to improve the mechanical stabilization of the fracture. However, its clinical use remains controversial partly because of concerns about implant loosening as well as frictional or constrictive interference between the cerclage and cortical surface. The aim of this study was to investigate in a distal tibia fracture model possible loosening of the cerclage and interference with the bone surface. Fourteen distal tibia oblique fractures (AO/OTA 42-A2) in seven human and synthetic bone samples, respectively were treated by plate osteosynthesis with supplemental cerclage wiring. The samples were subjected to 50,000 load cycles under combined axial and torsional loads. Angular and axial displacement were continuously monitored to identify possible cerclage loosening cerclage. Potential incision of the cerclage was inspected visually and on CT scans. Angular displacement was significantly influenced by the bone material (p ≤ 0.001) and type of osteosynthesis (p ≤ 0.001), while axial displacement was only influenced by the type of osteosynthesis (p ≤ 0.001). Lowest displacements were found in samples with plate and cable cerclage. Cerclage stability was maintained throughout the entire test period of 50,000 load cycles for human and synthetic samples. CT images revealed notches from the cerclage clamping mechanism but no incisions of the cable itself. In an oblique distal tibia fracture model, loosening of the cable cerclage was not detected. Full weight-bearing cyclic loads resulted in slight imprints of the cerclage crimp in both, human and synthetic samples. Following the surgical guidelines for careful cerclage installation, a supplemental cable cerclage has the potential to significantly increase the construct stability.

European Journal of Trauma and Emergency SurgeryVol. 51(1)
Paracelsus Medical University (AT), University Hospital Augsburg (DE), Berufsgenossenschaftliche Unfallklinik Murnau (DE)
Paracelsus Medizinische Privatuniversität
Openalex Percentile: Top 81%
Bone fractures and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.