Assessing the Impact of End-of-Life Processing on Reusable Medical Devices

End-of-life (EOL) testing requirements for reusable medical devices continue to cause confusion in the medical device industry. Regulatory expectations from U.S. and European Union authorities differ, particularly regarding whether manufacturers must conduct EOL testing or if a cautionary statement in the instructions for use is sufficient. ISO 10993-1:2021 mandates biological safety evaluations for the maximum validated number of processing cycles. Similarly, ISO 17664-1:2021, regulation (EU) 2017/745, and Food and Drug Administration guidance require manufacturers to determine whether repeated processing causes degradation that could limit a device's usable life. However, these documents provide limited direction on how to perform such assessments, leaving a gap in standardized methodology. This study aimed to develop a practical and reproducible protocol for EOL evaluation, reflecting real-world clinical use and manufacturer-recommended processing instructions. The primary objective was to assess whether biological or chemical residues remained on devices after repeated use and processing. A theoretical estimate of 100 use and processing cycles was used, representing a reasonable service life based on routine clinical conditions. A simulated 100-cycle protocol, including soiling, cleaning, and sterilization testing, revealed negligible to nondetectable residual contaminants on both surrogate coupons and actual devices. These results suggest that even under worst-case conditions, devices can be processed effectively without substantial accumulation of contaminants, supporting the selected EOL threshold. This approach offers a reproducible framework for manufacturers to evaluate and validate EOL claims, addressing a critical regulatory and practical challenge.

Authors

Publication Details

Journal
Biomedical Instrumentation & Technology
Published
2025-04-01
DOI
https://doi.org/10.2345/0899-8205-59.1.128
Primary Topic
Quality and Safety in Healthcare
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing the Impact of End-of-Life Processing on Reusable Medical Devices

Daniel B. Olsen, Alpa V. Patel
Biomedical Instrumentation & Technology
Quality and Safety in Healthcare
article

Assessing the Impact of End-of-Life Processing on Reusable Medical Devices

Daniel B. Olsen, Alpa V. Patel
article en

Abstract

End-of-life (EOL) testing requirements for reusable medical devices continue to cause confusion in the medical device industry. Regulatory expectations from U.S. and European Union authorities differ, particularly regarding whether manufacturers must conduct EOL testing or if a cautionary statement in the instructions for use is sufficient. ISO 10993-1:2021 mandates biological safety evaluations for the maximum validated number of processing cycles. Similarly, ISO 17664-1:2021, regulation (EU) 2017/745, and Food and Drug Administration guidance require manufacturers to determine whether repeated processing causes degradation that could limit a device's usable life. However, these documents provide limited direction on how to perform such assessments, leaving a gap in standardized methodology. This study aimed to develop a practical and reproducible protocol for EOL evaluation, reflecting real-world clinical use and manufacturer-recommended processing instructions. The primary objective was to assess whether biological or chemical residues remained on devices after repeated use and processing. A theoretical estimate of 100 use and processing cycles was used, representing a reasonable service life based on routine clinical conditions. A simulated 100-cycle protocol, including soiling, cleaning, and sterilization testing, revealed negligible to nondetectable residual contaminants on both surrogate coupons and actual devices. These results suggest that even under worst-case conditions, devices can be processed effectively without substantial accumulation of contaminants, supporting the selected EOL threshold. This approach offers a reproducible framework for manufacturers to evaluate and validate EOL claims, addressing a critical regulatory and practical challenge.

Biomedical Instrumentation & TechnologyVol. 59(1)
Openalex Percentile: Top 75%
Quality and Safety in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.