The Utility of a Modified Data-Driven Motion Correction Method Incorporating Computed Tomography Positional Data for Positron Emission Tomography

In positron emission tomography/computed tomography (PET/CT), long image acquisition time often causes patient motion and misalignment between the PET and CT images.This misalignment can compromise the accuracy of attenuation correction and quantitative values, thus affecting diagnostic reliability.Recently, data-driven motion correction (DDMC) methods have been developed to detect motion using brief static PET images, and our previous study also demonstrated the effectiveness of the DDMC method.However, DDMC assumes alignment between PET reference images and CT scans used for attenuation correction.When patients move after CT scan, the PET and CT images misalign, necessitating rescanning or manual realignment.The present study introduces a modified approach, namely DDMC+CT, that integrates CT positional data into the DDMC process to address PET/CT misalignment.We evaluated the DDMC+CT approach using a phantom study and clinical images.The phantom study simulated PET/CT misalignment by moving a Hoffman phantom post-CT scan, with contrast and standardized uptake values (SUV) compared with and without DDMC+CT.The clinical images included whole-body PET/CT scans with head and neck motion, where quantitative and visual assessments were performed.The DDMC+CT method successfully corrected PET/CT misalignment in phantom and clinical images, yielding SUV and contrast values consistent with reference images, resulting in improved diagnostic accuracy.The DDMC+CT method demonstrates a potential forA c c e p t e d M a n u s c r i p t clinical application, where it may minimize rescans and reduce patient and operator burden by allowing automatic misalignment correction.

Authors

Institutions

Publication Details

Journal
The Tohoku Journal of Experimental Medicine
Published
2025-01-01
DOI
https://doi.org/10.1620/tjem.2025.j072
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Utility of a Modified Data-Driven Motion Correction Method Incorporating Computed Tomography Positional Data for Positron Emission Tomography

Hiroshi Watabe, Akihito Usui, Kentaro Takanami, Hirofumi Kawakami et al.
The Tohoku Journal of Experimental Medicine
Medical Imaging Techniques and Applications
article

The Utility of a Modified Data-Driven Motion Correction Method Incorporating Computed Tomography Positional Data for Positron Emission Tomography

Hiroshi Watabe, Akihito Usui, Kentaro Takanami, Hirofumi Kawakami, Hiroyasu Kodama, Tatsuo Takeuchi, Tomohiro Kaneta, Yoshitaka Tanaka, Kei Takase, Hayato Odagiri, Akie Katsuki, Nobuo Yamaguchi
article en

Abstract

In positron emission tomography/computed tomography (PET/CT), long image acquisition time often causes patient motion and misalignment between the PET and CT images.This misalignment can compromise the accuracy of attenuation correction and quantitative values, thus affecting diagnostic reliability.Recently, data-driven motion correction (DDMC) methods have been developed to detect motion using brief static PET images, and our previous study also demonstrated the effectiveness of the DDMC method.However, DDMC assumes alignment between PET reference images and CT scans used for attenuation correction.When patients move after CT scan, the PET and CT images misalign, necessitating rescanning or manual realignment.The present study introduces a modified approach, namely DDMC+CT, that integrates CT positional data into the DDMC process to address PET/CT misalignment.We evaluated the DDMC+CT approach using a phantom study and clinical images.The phantom study simulated PET/CT misalignment by moving a Hoffman phantom post-CT scan, with contrast and standardized uptake values (SUV) compared with and without DDMC+CT.The clinical images included whole-body PET/CT scans with head and neck motion, where quantitative and visual assessments were performed.The DDMC+CT method successfully corrected PET/CT misalignment in phantom and clinical images, yielding SUV and contrast values consistent with reference images, resulting in improved diagnostic accuracy.The DDMC+CT method demonstrates a potential forA c c e p t e d M a n u s c r i p t clinical application, where it may minimize rescans and reduce patient and operator burden by allowing automatic misalignment correction.

The Tohoku Journal of Experimental Medicine
General Electric (Spain) (ES), Tohoku University (JP), Japan Radioisotope Association (JP), Tohoku University Hospital (JP), Cyclotron (Netherlands) (NL)
Openalex Percentile: Top 80%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.