Exposure control at the first dark-field chest radiography demonstrator system
An exposure planning approach was successfully implemented at the first clinical dark-field chest radiography system, delivering a detector dose within the required range for different patient sizes. A conventional radiograph with AEC is necessary for this implementation. The strong correlation between BMI and tube current offers an alternative approach for patient-specific exposure control.
Authors
- Theresa Urban (ORCID: https://orcid.org/0000-0002-0352-8180)
- Fabio De Marco (ORCID: https://orcid.org/0000-0002-3561-7305)
- M E Lochschmidt
- Henriette Bast (ORCID: https://orcid.org/0009-0001-7390-1809)
- Manuela Frank (ORCID: https://orcid.org/0000-0001-5642-0694)
- Daniela Pfeiffer (ORCID: https://orcid.org/0000-0003-2002-8337)
- Thomas Koehler (ORCID: https://orcid.org/0000-0002-1986-0057)
- R. Schick
- Wolfgang Noichl (ORCID: https://orcid.org/0000-0002-3588-8164)
- Ingo Maack
- Lennard Kaster
- K Willer
- Bernhard Renger (ORCID: https://orcid.org/0000-0001-8400-2377)
- G S Zimmermann
Institutions
- TUM Klinikum (DE)
- Philips (Germany) (DE)
- Technical University of Munich (DE)
Publication Details
- Journal
- Physica Medica
- Published
- 2025-06-11
- DOI
- https://doi.org/10.1016/j.ejmp.2025.105013
- Citations
- 3
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 2.02
Funders
- Institute for Advanced Study
- Karlsruhe Institute of Technology
- Australian Government
- Deutsche Forschungsgemeinschaft
- Bundesministerium für Bildung und Forschung
- Technische Universität München
- European Research Council
- Institute for Advanced Study, Technische Universität München
- H2020 European Research Council