Comprehensive Stiffness Modeling and Evaluation of an Orthopedic Surgical Robot for Enhanced Cutting Operation Performance

This study presents an integrated stiffness modeling and evaluation framework for an orthopedic surgical robot, aiming to enhance cutting accuracy and operational stability. A comprehensive stiffness model is developed, incorporating the stiffness of the end-effector, cutting tool, and force sensor. End-effector stiffness is computed using the virtual joint method based on the Jacobian matrix, enabling accurate analysis of stiffness distribution within the robot's workspace. Joint stiffness is experimentally identified through laser tracker-based displacement measurements under controlled loads and calculated using a least-squares method. The results show displacement errors below 0.3 mm and joint stiffness estimation errors under 1.5%, with values more consistent and stable than those reported for typical surgical robots. Simulation studies reveal spatial variations in operational stiffness, identifying zones of low stiffness and excessive stiffness. Compared to prior studies where stiffness varied over 50%, the proposed model exhibits superior uniformity. Experimental validation confirms model fidelity, with prediction errors generally below 5%. Cutting experiments on porcine femurs demonstrate real-world applicability, achieving average stiffness prediction errors below 3%, and under 1% in key directions. The model supports stiffness-aware trajectory planning and control, reducing cutting deviation by up to 10% and improving workspace stiffness stability by 30%. This research offers a validated, high-accuracy approach to stiffness modeling for surgical robots, bridging the gap between simulation and clinical application, and providing a foundation for safer, more precise robotic orthopedic procedures.

Authors

Institutions

Publication Details

Journal
Biomimetics
Published
2025-06-08
DOI
https://doi.org/10.3390/biomimetics10060383
Primary Topic
Robotic Mechanisms and Dynamics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comprehensive Stiffness Modeling and Evaluation of an Orthopedic Surgical Robot for Enhanced Cutting Operation Performance

Ji Tan, Guangqing Chen, Heqiang Tian, Zhuo Chen et al.
Biomimetics
Robotic Mechanisms and Dynamics
article

Comprehensive Stiffness Modeling and Evaluation of an Orthopedic Surgical Robot for Enhanced Cutting Operation Performance

Ji Tan, Guangqing Chen, Heqiang Tian, Zhuo Chen, Mengke Zhang
article en

Abstract

This study presents an integrated stiffness modeling and evaluation framework for an orthopedic surgical robot, aiming to enhance cutting accuracy and operational stability. A comprehensive stiffness model is developed, incorporating the stiffness of the end-effector, cutting tool, and force sensor. End-effector stiffness is computed using the virtual joint method based on the Jacobian matrix, enabling accurate analysis of stiffness distribution within the robot's workspace. Joint stiffness is experimentally identified through laser tracker-based displacement measurements under controlled loads and calculated using a least-squares method. The results show displacement errors below 0.3 mm and joint stiffness estimation errors under 1.5%, with values more consistent and stable than those reported for typical surgical robots. Simulation studies reveal spatial variations in operational stiffness, identifying zones of low stiffness and excessive stiffness. Compared to prior studies where stiffness varied over 50%, the proposed model exhibits superior uniformity. Experimental validation confirms model fidelity, with prediction errors generally below 5%. Cutting experiments on porcine femurs demonstrate real-world applicability, achieving average stiffness prediction errors below 3%, and under 1% in key directions. The model supports stiffness-aware trajectory planning and control, reducing cutting deviation by up to 10% and improving workspace stiffness stability by 30%. This research offers a validated, high-accuracy approach to stiffness modeling for surgical robots, bridging the gap between simulation and clinical application, and providing a foundation for safer, more precise robotic orthopedic procedures.

BiomimeticsVol. 10(6)
Shandong University of Science and Technology (CN)
National Natural Science Foundation of China, Natural Science Foundation of Shandong Province
Openalex Percentile: Top 86%
Robotic Mechanisms and Dynamics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.