Simulation and one-ring prototyping of 1 mm-rod-resolution hemispherical brain PET with TOF-DOI detectors
<i>Objective.</i>Brain positron emission tomography (PET) imaging plays crucial roles in research and diagnosis of various brain diseases. To achieve high spatial resolution and high sensitivity, we proposed a hemispherical geometry which offers higher sensitivity with fewer detectors than a conventional cylindrical geometry. Our developed hemispherical brain PET system, Vrain, has indeed achieved a rod resolution of 2.2 mm with a 229 ps time-of-flight (TOF) resolution. To further improve the spatial resolution, we will use TOF and depth-of-interaction (DOI) detectors with our original crosshair light-sharing (CLS) configuration. This study aimed at estimating the performance of the hemispherical brain PET with TOF-DOI detectors and at developing a one-ring PET prototype with 1.6 mm scintillator pitch CLS-based TOF-DOI detectors.<i>Approach.</i>The sensitivity, rod resolution, and image quality of the TOF-DOI hemispherical brain PET (TDHBP-sim) and Vrain (Vrain-sim) were estimated using Geant4 simulation. A one-ring prototype with a 30 cm diameter was developed using the CLS-based TOF-DOI detectors. The energy resolution, TOF timing resolution, rod resolution, and the Hoffman brain phantom image quality of the prototype were evaluated.<i>Main results.</i>In the simulation study, TDHBP-sim achieved 1.4 times better sensitivity than Vrain-sim. TDHBP-sim visualized 1.0 mm rods and gyri and sulci structures in the brain phantom. In the one-ring experiment, the energy resolution was 11.6% at 511 keV, the TOF timing resolution was 294.6 ps, and 1.0 mm rods were resolved at the central 10 cm-diameter field-of-view. The 0.8 mm-thick radioactivity distribution could be identified in the Hoffman phantom.<i>Significance.</i>The study findings suggested that a hemispherical brain PET with 1.6 mm scintillator pitch TOF-DOI detectors should offer excellent performance including 1 mm rod resolution.
Authors
- Hideaki Tashima (ORCID: https://orcid.org/0000-0003-3155-0083)
- Taiga Yamaya
- Go Akamatsu (ORCID: https://orcid.org/0000-0001-9686-8901)
- Eiji Yoshida (ORCID: https://orcid.org/0000-0003-4269-2618)
- M. Takahashi (ORCID: https://orcid.org/0000-0002-1953-7428)
- Yuma Iwao
- K Narita
Institutions
- Chiba University (JP)
- National Institutes for Quantum Science and Technology (JP)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2025-06-09
- DOI
- https://doi.org/10.1088/1361-6560/ade2b4
- Citations
- 2
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 2.72
Funders
- Japan Agency for Medical Research and Development