Modelling and numerical solution of variable orifice flow meter for application of respiratory

This study investigates three-dimensional computational fluid dynamics (CFD) analysis related to Variable Orifice Flowmeters (VOFM) used in mechanical ventilators. The analysis employed a dynamic fluid-structure interaction (FSI) approach, utilizing the finite volume method (FVM). A comprehensive numerical simulation was performed using two turbulence k-ω SST model. The CFD methodology was validated by examining the flow characteristics of flexible membranes making triangular and circular orifice plates. Validation was achieved by comparing the numerical results with previously published experimental data, revealing several high correlation factors that describe the deflection of the flexible membrane, drag coefficient, and pressure drop in relation to the mass flowrate. The findings indicate that both the thickness and shape of the orifice play a crucial role in influencing pressure drop and deflection. Notably, the circular model exhibits more linear behaviour and greater sensitivity compared to the triangular model.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2025-06-07
DOI
https://doi.org/10.1016/j.medengphy.2025.104379
Citations
1
Primary Topic
Flow Measurement and Analysis
Type
article
Field-Weighted Citation Impact
0.86
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Modelling and numerical solution of variable orifice flow meter for application of respiratory

Muneer A. Ismael, Rana K. Shamkhi
1 citations
Medical Engineering & Physics
Flow Measurement and Analysis
0.86
article

Modelling and numerical solution of variable orifice flow meter for application of respiratory

Muneer A. Ismael, Rana K. Shamkhi
article en
1 citations

Abstract

This study investigates three-dimensional computational fluid dynamics (CFD) analysis related to Variable Orifice Flowmeters (VOFM) used in mechanical ventilators. The analysis employed a dynamic fluid-structure interaction (FSI) approach, utilizing the finite volume method (FVM). A comprehensive numerical simulation was performed using two turbulence k-ω SST model. The CFD methodology was validated by examining the flow characteristics of flexible membranes making triangular and circular orifice plates. Validation was achieved by comparing the numerical results with previously published experimental data, revealing several high correlation factors that describe the deflection of the flexible membrane, drag coefficient, and pressure drop in relation to the mass flowrate. The findings indicate that both the thickness and shape of the orifice play a crucial role in influencing pressure drop and deflection. Notably, the circular model exhibits more linear behaviour and greater sensitivity compared to the triangular model.

Medical Engineering & PhysicsVol. 142(1)
Iraq University College (IQ)
Openalex Percentile: Top 30%
Flow Measurement and Analysis
0.86
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Modelling and numerical solution of variable orifice flow meter for application of respiratory — Muneer A. Ismael, Rana K. Shamkhi · Medical Engineering & Physics (2025) | TGRS Research Map | TGRS