AN OPTICAL METHOD FOR CORRECTING THE EMISSIVITY OF BLACKBODY MODELS

Modern infrared thermometers are designed to measure the temperature of objects with an emissivity from 0.1 to 1.0. However, when approving the type, checking or calibrating infrared thermometers, metrological characteristics are determined only when the emissivity is close to one. It is impossible to perform objective control at other emissivity values due to the lack of necessary standards. The paper describes an optical method for correcting the emissivity of black body models. The method is based on the use of the properties of reflection and refraction of light at homogeneous flat interfaces of media. Viewing windows made of optical glass of the KI brand, calcium fluoride of the FC-I brand and potassium bromide are used as media separators. The possibility of reducing the emissivity of a black body to ≈ (0.5…0.6) is theoretically justified and experimentally confirmed. The additional temperature reproduction error, depending on the temperature and the material of the viewing windows, does not exceed 2.0…4.7 °C. The optical method of emissivity correction can be used to: improve reference models of a black body; calibrate infrared thermometers with an established emissivity of less than 0.99; verify algorithms for accounting for the emissivity of an object implemented in verifiable pyrometers.

Authors

Publication Details

Journal
Kontrol Diagnostika
Published
2025-03-01
DOI
https://doi.org/10.14489/td.2025.03.pp.044-055
Primary Topic
Remote Sensing in Agriculture
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

AN OPTICAL METHOD FOR CORRECTING THE EMISSIVITY OF BLACKBODY MODELS

M. V. Golobokov
Kontrol Diagnostika
Remote Sensing in Agriculture
article

AN OPTICAL METHOD FOR CORRECTING THE EMISSIVITY OF BLACKBODY MODELS

M. V. Golobokov
article en

Abstract

Modern infrared thermometers are designed to measure the temperature of objects with an emissivity from 0.1 to 1.0. However, when approving the type, checking or calibrating infrared thermometers, metrological characteristics are determined only when the emissivity is close to one. It is impossible to perform objective control at other emissivity values due to the lack of necessary standards. The paper describes an optical method for correcting the emissivity of black body models. The method is based on the use of the properties of reflection and refraction of light at homogeneous flat interfaces of media. Viewing windows made of optical glass of the KI brand, calcium fluoride of the FC-I brand and potassium bromide are used as media separators. The possibility of reducing the emissivity of a black body to ≈ (0.5…0.6) is theoretically justified and experimentally confirmed. The additional temperature reproduction error, depending on the temperature and the material of the viewing windows, does not exceed 2.0…4.7 °C. The optical method of emissivity correction can be used to: improve reference models of a black body; calibrate infrared thermometers with an established emissivity of less than 0.99; verify algorithms for accounting for the emissivity of an object implemented in verifiable pyrometers.

Kontrol Diagnostika(321)
Openalex Percentile: Top 87%
Remote Sensing in Agriculture
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.