Reporting Clinical Results of EPID-Based Patient-Specific Quality Assurance for Brain Stereotactic Radiosurgery/Radiation Therapy: Demonstrating the Feasibility of Same-Day Stereotactic Radiosurgery via HyperArc Delivery
These promising PSQA results for complex clinical HyperArc brain SRS/SRT plans delivered in a timely manner via EPID-based PD and independent MC second check demonstrate EPID-based PD is a fast, safe, and efficient QA method for HyperArc treatments. With proper automation of contouring tool and SRS planning via RapidPlan model, this EPID-based PSQA method can be safely implemented for the same-day HyperArc delivery of high-quality brain SRS/SRT treatments in the near future.
Authors
- William St Clair
- Shane McCarthy (ORCID: https://orcid.org/0000-0002-2715-4187)
- Damodar Pokhrel (ORCID: https://orcid.org/0000-0003-2694-1047)
- Joshua Misa (ORCID: https://orcid.org/0009-0007-8084-0438)
Institutions
- University of Kentucky (US)
Publication Details
- Journal
- Advances in Radiation Oncology
- Published
- 2025-06-07
- DOI
- https://doi.org/10.1016/j.adro.2025.101831
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Varian Medical Systems