Validation of real-time aging simulation of poly(lactic acid) (PLA) using accelerated aging in accordance with ASTM F1980

To advance the application of bio-based materials in medical technology, further research is required to assess their long-term performance. In common practice, accelerated aging tests based on ASTM F1980 are used in medical contexts to predict material behavior over time. This standard provides calculation guidelines to determine the equivalent storage duration under artificially accelerated aging conditions using a Q 10 -factor. This factor, typically assumed to be 2, represents the increase in reaction rate due to elevated temperatures. In this study, a comparison was conducted between accelerated aged samples and their real-time equivalents using various PLA types. The results indicate that the standard assumption of Q 10 = 2 can lead to an overestimation of degradation, resulting in a misrepresentation of real-time aging behavior. This discrepancy is substantiated by experimental data, including mechanical, thermal, and chemical analyses. A key factor contributing to this deviation appears to be the reliance on overly simplistic assumptions regarding degradation kinetics, which fail to account for autocatalytic reactions and the inherently multi-stage nature of the degradation process. In the present study, this observation was further corroborated through the determination of material-specific Q 10 -factors. These factors, found to range between 2.3 and 2.5, exhibited dynamic variations throughout the degradation process, highlighting the need for a refined approach to accelerated aging methodologies. • Accelerated aging tests are the preferred method for assessing long-term behavior • Excessive moisture uptake accelerates hydrolytic degradation during aging tests • Conventional Q 10 -factor of 2 overestimates real-time aging effects of PLA • Using a material-specific Q 10 -factor improves the accuracy of aging simulations • Q 10 = 2.5 reduces the required storage time for one year from 91.3 to 58.4 days

Authors

Institutions

Publication Details

Journal
Polymer Testing
Published
2025-06-06
DOI
https://doi.org/10.1016/j.polymertesting.2025.108880
Citations
4
Primary Topic
biodegradable polymer synthesis and properties
Type
article
Field-Weighted Citation Impact
2.39
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of real-time aging simulation of poly(lactic acid) (PLA) using accelerated aging in accordance with ASTM F1980

Jan‐Christoph Zarges, Hans‐Peter Heim, N. Krug, Chaimae Chettouh
4 citations
Polymer Testing
biodegradable polymer synthesis and properties
2.39
article

Validation of real-time aging simulation of poly(lactic acid) (PLA) using accelerated aging in accordance with ASTM F1980

Jan‐Christoph Zarges, Hans‐Peter Heim, N. Krug, Chaimae Chettouh
article en
4 citations

Abstract

To advance the application of bio-based materials in medical technology, further research is required to assess their long-term performance. In common practice, accelerated aging tests based on ASTM F1980 are used in medical contexts to predict material behavior over time. This standard provides calculation guidelines to determine the equivalent storage duration under artificially accelerated aging conditions using a Q 10 -factor. This factor, typically assumed to be 2, represents the increase in reaction rate due to elevated temperatures. In this study, a comparison was conducted between accelerated aged samples and their real-time equivalents using various PLA types. The results indicate that the standard assumption of Q 10 = 2 can lead to an overestimation of degradation, resulting in a misrepresentation of real-time aging behavior. This discrepancy is substantiated by experimental data, including mechanical, thermal, and chemical analyses. A key factor contributing to this deviation appears to be the reliance on overly simplistic assumptions regarding degradation kinetics, which fail to account for autocatalytic reactions and the inherently multi-stage nature of the degradation process. In the present study, this observation was further corroborated through the determination of material-specific Q 10 -factors. These factors, found to range between 2.3 and 2.5, exhibited dynamic variations throughout the degradation process, highlighting the need for a refined approach to accelerated aging methodologies. • Accelerated aging tests are the preferred method for assessing long-term behavior • Excessive moisture uptake accelerates hydrolytic degradation during aging tests • Conventional Q 10 -factor of 2 overestimates real-time aging effects of PLA • Using a material-specific Q 10 -factor improves the accuracy of aging simulations • Q 10 = 2.5 reduces the required storage time for one year from 91.3 to 58.4 days

Polymer TestingVol. 149
University of Kassel (DE)
Openalex Percentile: Top 12%
biodegradable polymer synthesis and properties
2.39
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.