Abstract The X-ray mirror surface requires high spatial resolution image measurement, a task for which the white light interferometer stitching method is most commonly used in the X-ray optical measurement lab. The measurement stage design, a crucial part of this method, includes two linear axes and two rotation axes, each with specific travel ranges. The 1200 mm travel range axis is the X axis, and the 390 mm travel range axis is the Y axis. The measurement head travel range is 390 mm along the Z axis. The rotation axes are the X-axis and the Y-axis, and each rotation range is ± 1.5 degrees. The accuracy of the rotation axes is 2.87 x10 −3 degrees. The linear axes stage X, Y, and Z-axis scales accuracy ±1 µm for up to 1 m. In this study, the measurement head is ZYGO Nexview NX2, which has one rotation axis to the Y-axis. The Nexview NX2 measurement sensor array pixels are 1000 * 1000 and 1X objective head with an 8.69 * 8.69 mm field of view. Thus, each pixel’s spatial resolution is about 8.69 μm for the white light interferometer stitching measurement with 1X objective. Therefore, the stage linear direction accuracy is good enough to support the different snapshot images for stitching.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2025-05-01
DOI
https://doi.org/10.1088/1742-6596/3010/1/012088
Primary Topic
Advanced X-ray Imaging Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The white light interferometer stitching measurement stage design for the X-ray mirror

Gung-Chian Yin, Ming-Ying Hsu, Chien‐Chung Fu, Hok‐Sum Fung et al.
Journal of Physics Conference Series
Advanced X-ray Imaging Techniques
article

The white light interferometer stitching measurement stage design for the X-ray mirror

Gung-Chian Yin, Ming-Ying Hsu, Chien‐Chung Fu, Hok‐Sum Fung, Shang-Wei Lin, Bo-Yi Chen
article en

Abstract

Abstract The X-ray mirror surface requires high spatial resolution image measurement, a task for which the white light interferometer stitching method is most commonly used in the X-ray optical measurement lab. The measurement stage design, a crucial part of this method, includes two linear axes and two rotation axes, each with specific travel ranges. The 1200 mm travel range axis is the X axis, and the 390 mm travel range axis is the Y axis. The measurement head travel range is 390 mm along the Z axis. The rotation axes are the X-axis and the Y-axis, and each rotation range is ± 1.5 degrees. The accuracy of the rotation axes is 2.87 x10 −3 degrees. The linear axes stage X, Y, and Z-axis scales accuracy ±1 µm for up to 1 m. In this study, the measurement head is ZYGO Nexview NX2, which has one rotation axis to the Y-axis. The Nexview NX2 measurement sensor array pixels are 1000 * 1000 and 1X objective head with an 8.69 * 8.69 mm field of view. Thus, each pixel’s spatial resolution is about 8.69 μm for the white light interferometer stitching measurement with 1X objective. Therefore, the stage linear direction accuracy is good enough to support the different snapshot images for stitching.

Journal of Physics Conference SeriesVol. 3010(1)
National Synchrotron Radiation Research Center (TW)
Openalex Percentile: Top 82%
Advanced X-ray Imaging Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.