Advanced X-ray PIxel Detector (AXPiDe v2.0): new modular multichannel detector based on SDD available at the XAFS beamline of Elettra

Abstract This contribution will report on a detection system specially designed and developed to fulfil the needs of X-ray Absorption Spectroscopy (XAS) experiments at the XAFS beamline of the ELETTRA synchrotron. It is composed of 8 monolithic multipixel arrays of Silicon Drift Detectors (SDDs), each comprising 8 cells (3x3 mm 2 ) fabricated on 450- µ m-thick, n-type, high-purity silicon wafers, and it includes a Tungsten collimation system. This results in 64 independent cells for a total collimated area of 500 mm 2 . All arrays are connected to separate back-end electronics, and they are calibrated, aligned and summed together by the acquisition software. The system includes custom-made, ultra-low-noise front-end electronics, a dedicated acquisition system, digital filtering, and temperature control and stabilization. The sensor is optimized to operate in the energy range 3-30 keV. A dedicated acquisition software, Fluorescence Instrumentation Control Universal Software (FICUS), developed using NI LabVIEW and fully integrated with the control system of the beamline, allows the instrument performance to be controlled, fine-tuned and monitored. Accurate characterization performed at room temperature at the XAFS beamline in Elettra demonstrated very interesting results in terms of energy resolution, with a FWHM below 170 eV at the 5.9 keV Mn-K α line for the sum of all cells, a high count rate capability with an excellent peak-to-background ratio. All these specifications make it possible to collect high-data-quality XAS spectra on diluted elements embedded in heavy matrices, which can be exploited to improve the throughput of the beamline, as well as to follow slow kinetics.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2025-05-01
DOI
https://doi.org/10.1088/1742-6596/3010/1/012142
Citations
1
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
1.87
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Advanced X-ray PIxel Detector (AXPiDe v2.0): new modular multichannel detector based on SDD available at the XAFS beamline of Elettra

Luca Olivi, L. Stebel, G. Orzan, A. Vacchi et al.
1 citations
Journal of Physics Conference Series
Particle Detector Development and Performance
1.87
article

Advanced X-ray PIxel Detector (AXPiDe v2.0): new modular multichannel detector based on SDD available at the XAFS beamline of Elettra

Luca Olivi, L. Stebel, G. Orzan, A. Vacchi, D. Giuressi, Filippo Mele, A. Picciotto, F. Ficorella, R.H. Menk, G. Zampa, Giuliana Aquilanti, G. Cautero, D. Cirrincione, M. Antonelli, G. Bertuccio, G. Pepponi, Giovanni Agostini, A. Rachevski, N. Zampa, I. Rashevskaya, N. Zorzi
article en
1 citations

Abstract

Abstract This contribution will report on a detection system specially designed and developed to fulfil the needs of X-ray Absorption Spectroscopy (XAS) experiments at the XAFS beamline of the ELETTRA synchrotron. It is composed of 8 monolithic multipixel arrays of Silicon Drift Detectors (SDDs), each comprising 8 cells (3x3 mm 2 ) fabricated on 450- µ m-thick, n-type, high-purity silicon wafers, and it includes a Tungsten collimation system. This results in 64 independent cells for a total collimated area of 500 mm 2 . All arrays are connected to separate back-end electronics, and they are calibrated, aligned and summed together by the acquisition software. The system includes custom-made, ultra-low-noise front-end electronics, a dedicated acquisition system, digital filtering, and temperature control and stabilization. The sensor is optimized to operate in the energy range 3-30 keV. A dedicated acquisition software, Fluorescence Instrumentation Control Universal Software (FICUS), developed using NI LabVIEW and fully integrated with the control system of the beamline, allows the instrument performance to be controlled, fine-tuned and monitored. Accurate characterization performed at room temperature at the XAFS beamline in Elettra demonstrated very interesting results in terms of energy resolution, with a FWHM below 170 eV at the 5.9 keV Mn-K α line for the sum of all cells, a high count rate capability with an excellent peak-to-background ratio. All these specifications make it possible to collect high-data-quality XAS spectra on diluted elements embedded in heavy matrices, which can be exploited to improve the throughput of the beamline, as well as to follow slow kinetics.

Journal of Physics Conference SeriesVol. 3010(1)
University of Udine (IT), Fondazione Bruno Kessler (IT), Elettra-Sincrotrone Trieste S.C.p.A. (IT), Istituto Nazionale di Fisica Nucleare, Trento Institute for Fundamental Physics And Applications (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Milano (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Trieste (IT), Mid Sweden University (SE), Politecnico di Milano (IT)
Openalex Percentile: Top 14%
Particle Detector Development and Performance
1.87
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.