Intra and inter-system reliability of a markerless motion capture system in different configurations during functional tasks: Implications for utility
Larger field of view, suboptimal camera placement and tasks performed away from the center of the field of view presented challenges to the DARI markerless mocap system. Intra-device and inter-device reliability can achieve a level of ICC > 0.7 using DARI markerless mocap technology using an optimized setup, but the transverse plane kinematics exhibit the weakest performance. Optimal camera placement (i.e., field of view) is critical for reliable results.