Effect of inaccurate b‐values from imaging gradients on intravoxel incoherent motion
Ignoring imaging gradients in b-value calculations introduces unnecessary inaccuracies, making IVIM results spurious and highly dependent on specific pulse sequence design and imaging parameters. These inaccuracies can be corrected by adjusting the b-value calculations, without additional measurements.
Authors
- Lars Olsson (ORCID: https://orcid.org/0000-0001-6090-6697)
- Adalsteinn Gunnlaugsson (ORCID: https://orcid.org/0000-0003-1354-8537)
- Patrik Brynolfsson (ORCID: https://orcid.org/0000-0002-0455-8904)
- Filip Szczepankiewicz (ORCID: https://orcid.org/0000-0002-5251-587X)
- Ivan A. Rashid
Institutions
- Lund University (SE)
- Skåne University Hospital (SE)
Publication Details
- Journal
- Magnetic Resonance in Medicine
- Published
- 2025-06-02
- DOI
- https://doi.org/10.1002/mrm.30579
- Citations
- 2
- Primary Topic
- MRI in cancer diagnosis
- Type
- article
- Field-Weighted Citation Impact
- 2.72
Funders
- Cancerfonden
- Fru Berta Kamprads Stiftelse
- Allmänna Sjukhusets i Malmö Stiftelse för Bekämpande av Cancer