Validation of an MEMS-Based Pressure Sensor System for Atrial Fibrillation Detection From Wrist and Finger

To address the unmet need for a low-cost, low-power wearable solution for continuous cardiovascular health monitoring, we developed and validated an Atrial Fibrillation (AF) detection algorithm using clinical data collected with a microelectromechanical system (MEMS)-based pressure sensor. This sensor system, consisting of a circuit board, capacitive digitizer, and three MEMS elements, was specifically designed for early detection of AF—a common cardiac arrhythmia that requires frequent screening. The proposed algorithm extracts seven AF-related features, derived from autocorrelation analysis, Inter-Beat Interval (IBI) measurements, and differential IBI (dIBI) analysis, including a novel Mean Distance of Points in the Poincaré Plot (MDPP) feature. Clinical validation was conducted using data from 53 participants across three datasets: 13 healthy volunteers (wrist), 20 post-cardiac surgery sinus rhythm patients (wrist), and 20 AF patients (wrist and finger). Leave-one-out cross-validation showed that logistic regression achieved an AUROC of 93.0% using the full feature set. Performance remained stable across segment lengths ranging from 10 to 120 seconds, supporting the algorithm’s suitability for continuous monitoring. Consistent performance across seven different classifiers (average AUROC 92.1%) further demonstrated the clinical applicability and generalizability of the approach for wearable-based AF screening. To assess robustness against motion artifacts, we introduced five types of synthetic noise, with the algorithm maintaining strong AF detection performance under these conditions. Finally, a systematic evaluation of sensor waveform shape and signal strength across sinus rhythm and AF, at both wrist and finger sites demonstrates the potential of the sensor system for wearable AF screening.

Authors

Institutions

Publication Details

Journal
IEEE Sensors Journal
Published
2025-06-03
DOI
https://doi.org/10.1109/jsen.2025.3574232
Citations
1
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of an MEMS-Based Pressure Sensor System for Atrial Fibrillation Detection From Wrist and Finger

Juhani Airaksinen, Samuli Jaakkola, Matti Kaisti, Ismail Elnaggar et al.
1 citations
IEEE Sensors Journal
Advanced Sensor and Energy Harvesting Materials
0.67
article

Validation of an MEMS-Based Pressure Sensor System for Atrial Fibrillation Detection From Wrist and Finger

Juhani Airaksinen, Samuli Jaakkola, Matti Kaisti, Ismail Elnaggar, Tero Koivisto, Tuija Vasankari, Olli Lahdenoja, Tuomas Kiviniemi, Yingjiang Zhao
article en
1 citations

Abstract

To address the unmet need for a low-cost, low-power wearable solution for continuous cardiovascular health monitoring, we developed and validated an Atrial Fibrillation (AF) detection algorithm using clinical data collected with a microelectromechanical system (MEMS)-based pressure sensor. This sensor system, consisting of a circuit board, capacitive digitizer, and three MEMS elements, was specifically designed for early detection of AF—a common cardiac arrhythmia that requires frequent screening. The proposed algorithm extracts seven AF-related features, derived from autocorrelation analysis, Inter-Beat Interval (IBI) measurements, and differential IBI (dIBI) analysis, including a novel Mean Distance of Points in the Poincaré Plot (MDPP) feature. Clinical validation was conducted using data from 53 participants across three datasets: 13 healthy volunteers (wrist), 20 post-cardiac surgery sinus rhythm patients (wrist), and 20 AF patients (wrist and finger). Leave-one-out cross-validation showed that logistic regression achieved an AUROC of 93.0% using the full feature set. Performance remained stable across segment lengths ranging from 10 to 120 seconds, supporting the algorithm’s suitability for continuous monitoring. Consistent performance across seven different classifiers (average AUROC 92.1%) further demonstrated the clinical applicability and generalizability of the approach for wearable-based AF screening. To assess robustness against motion artifacts, we introduced five types of synthetic noise, with the algorithm maintaining strong AF detection performance under these conditions. Finally, a systematic evaluation of sensor waveform shape and signal strength across sinus rhythm and AF, at both wrist and finger sites demonstrates the potential of the sensor system for wearable AF screening.

IEEE Sensors JournalVol. 25(13)
University of Turku (FI), Turku University Hospital (FI)
Business Finland
Openalex Percentile: Top 31%
Advanced Sensor and Energy Harvesting Materials
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.