Application of high-performance liquid chromatography with electrochemical detection for the analysis of N-(1,3-dimethylbutyl)-N’-phenyl-p-phenylenediamine quinone in ambient particulates

• A novel HPLC-ECD method was established to detect 6-PPDQ in PM samples. • A dual-mode ECD and C30 column enabled the selective and sensitive detection of 6-PPDQ. • First application of electrochemical detection for airborne 6-PPDQ quantification • This method achieved a low LOQ suitable for trace-level analysis of PM-bound 6-PPDQ. A simple and sensitive high-performance liquid chromatography (HPLC) method with electrochemical detection (ECD) was developed for measuring N -(1,3-dimethylbutyl)- N’ -phenyl- p -phenylenediamine quinone (6-PPDQ) in environmental samples. The system used C30 columns for sample separation and a dual-mode electrochemical detector with a reduction cell, followed by an oxidation/detection cell connected in series. Under the optimal reduction voltage (–600 mV vs. Au) and optimal oxidation voltage (+600 mV vs. Ag/AgCl), the assay demonstrated high accuracy (96%–107%) when applied to airborne particulate sample extracts spiked with known amounts of 6-PPDQ. The detection limit was 3 fmol per injection (signal-to-noise ratio = 3), with a calibration range of 0.1–100 pmol, exhibiting excellent linearity ( R 2 > 0.9995). Using this newly developed HPLC-ECD analytical method with simple sample preparation, daily variations in 6-PPDQ concentrations in airborne particles collected in Kyoto, Japan, ranging from 21 to 165 fmol m –3 (n = 20), were successfully measured, highlighting the high potential of this HPLC-ECD method for routine trace analysis in environmental samples.

Authors

Institutions

Publication Details

Journal
Journal of Chromatography Open
Published
2025-05-01
DOI
https://doi.org/10.1016/j.jcoa.2025.100232
Primary Topic
Analytical Chemistry and Chromatography
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Application of high-performance liquid chromatography with electrochemical detection for the analysis of N-(1,3-dimethylbutyl)-N’-phenyl-p-phenylenediamine quinone in ambient particulates

Akira Itô, Takayuki Kameda, Nozomu Tsuchiya, Vonika Ka‐Man Au et al.
Journal of Chromatography Open
Analytical Chemistry and Chromatography
article

Application of high-performance liquid chromatography with electrochemical detection for the analysis of N-(1,3-dimethylbutyl)-N’-phenyl-p-phenylenediamine quinone in ambient particulates

Akira Itô, Takayuki Kameda, Nozomu Tsuchiya, Vonika Ka‐Man Au, Natsumi Okawa
article en

Abstract

• A novel HPLC-ECD method was established to detect 6-PPDQ in PM samples. • A dual-mode ECD and C30 column enabled the selective and sensitive detection of 6-PPDQ. • First application of electrochemical detection for airborne 6-PPDQ quantification • This method achieved a low LOQ suitable for trace-level analysis of PM-bound 6-PPDQ. A simple and sensitive high-performance liquid chromatography (HPLC) method with electrochemical detection (ECD) was developed for measuring N -(1,3-dimethylbutyl)- N’ -phenyl- p -phenylenediamine quinone (6-PPDQ) in environmental samples. The system used C30 columns for sample separation and a dual-mode electrochemical detector with a reduction cell, followed by an oxidation/detection cell connected in series. Under the optimal reduction voltage (–600 mV vs. Au) and optimal oxidation voltage (+600 mV vs. Ag/AgCl), the assay demonstrated high accuracy (96%–107%) when applied to airborne particulate sample extracts spiked with known amounts of 6-PPDQ. The detection limit was 3 fmol per injection (signal-to-noise ratio = 3), with a calibration range of 0.1–100 pmol, exhibiting excellent linearity ( R 2 > 0.9995). Using this newly developed HPLC-ECD analytical method with simple sample preparation, daily variations in 6-PPDQ concentrations in airborne particles collected in Kyoto, Japan, ranging from 21 to 165 fmol m –3 (n = 20), were successfully measured, highlighting the high potential of this HPLC-ECD method for routine trace analysis in environmental samples.

Journal of Chromatography OpenVol. 7
Kyoto University (JP), Kyoto University of Education (JP)
Kanazawa University
Openalex Percentile: Top 89%
Analytical Chemistry and Chromatography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Application of high-performance liquid chromatography with electrochemical detection for the analysis of N-(1,3-dimethylbutyl)-N’-phenyl-p-phenylenediamine quinone in ambient particulates — Akira Itô, Takayuki Kameda, et al. · Journal of Chromatography Open (2025) | TGRS Research Map | TGRS