Evaluation of Screw Loosening and Removal Force of Original and Compatible Implant-Abutment Connection (Tissue Level): In Vitro Study

Abstract Screw loosening in dental implant prosthetics can lead to serious issues. Various factors contribute to screw loosening, including prosthesis design, misalignment between the prosthesis and abutment or between the abutment and implant, insufficient screw tightening force, excessive biomechanical loading, the material and design of the screw, chewing forces, and parafunctional habits. This study evaluated three implant systems (SNUCone, Dentis, and Implant Swiss) along with their original and compatible abutments. The implants were mounted in acrylic blocks measuring 20 × 6 × 10 mm, and the abutments were connected to the fixtures with a torque of 30 N/cm. The samples were subjected to a chewing simulator with a load of 90 N and 10,000 cycles, and the amount of screw loosening was measured. Subsequently, the removal torque of the abutments was recorded using a universal testing machine. The data were analyzed using analysis of variance and Tukey's post hoc test. According to the study results, the screw loosening torque of abutments is significantly influenced by the type of abutment and implant. Among the groups with Implant Swiss fixtures, the Implant Swiss abutments had the highest loosening torque, with an average of 27.24 N/cm, whereas in SNUCone fixtures, Dentis abutments showed the lowest loosening torque. Similarly, in Implant Swiss fixtures, Dentis abutments recorded the lowest loosening torque, with an average of 23.86 N/cm. In SNUCone implants, Dentis abutments had the lowest loosening torque, averaging 23.86 N/cm. The removal torque required for abutments did not significantly differ across the different implant systems. According to the study findings, the screw loosening torque of abutments varies depending on the type of abutment and implant. Implant Swiss abutments exhibited the highest loosening torque, while SNUCone abutments showed the lowest. The highest average abutment removal torque was observed in the Implant Swiss implant with the SNUCone abutment, and the lowest was in the Dentis implant with the Implant Swiss abutment. The abutment type has a limited effect on this parameter, which may be influenced by other factors such as the overall design of the implant systems.

Authors

Institutions

Publication Details

Journal
European Dental Research and Biomaterials Journal
Published
2025-01-01
DOI
https://doi.org/10.1055/s-0045-1809346
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Screw Loosening and Removal Force of Original and Compatible Implant-Abutment Connection (Tissue Level): In Vitro Study

Ehsan Ghasemi, Morteza Ahmadi, Amir H. Fathi, Sakineh Hasani
European Dental Research and Biomaterials Journal
Dental Implant Techniques and Outcomes
article

Evaluation of Screw Loosening and Removal Force of Original and Compatible Implant-Abutment Connection (Tissue Level): In Vitro Study

Ehsan Ghasemi, Morteza Ahmadi, Amir H. Fathi, Sakineh Hasani
article en

Abstract

Abstract Screw loosening in dental implant prosthetics can lead to serious issues. Various factors contribute to screw loosening, including prosthesis design, misalignment between the prosthesis and abutment or between the abutment and implant, insufficient screw tightening force, excessive biomechanical loading, the material and design of the screw, chewing forces, and parafunctional habits. This study evaluated three implant systems (SNUCone, Dentis, and Implant Swiss) along with their original and compatible abutments. The implants were mounted in acrylic blocks measuring 20 × 6 × 10 mm, and the abutments were connected to the fixtures with a torque of 30 N/cm. The samples were subjected to a chewing simulator with a load of 90 N and 10,000 cycles, and the amount of screw loosening was measured. Subsequently, the removal torque of the abutments was recorded using a universal testing machine. The data were analyzed using analysis of variance and Tukey's post hoc test. According to the study results, the screw loosening torque of abutments is significantly influenced by the type of abutment and implant. Among the groups with Implant Swiss fixtures, the Implant Swiss abutments had the highest loosening torque, with an average of 27.24 N/cm, whereas in SNUCone fixtures, Dentis abutments showed the lowest loosening torque. Similarly, in Implant Swiss fixtures, Dentis abutments recorded the lowest loosening torque, with an average of 23.86 N/cm. In SNUCone implants, Dentis abutments had the lowest loosening torque, averaging 23.86 N/cm. The removal torque required for abutments did not significantly differ across the different implant systems. According to the study findings, the screw loosening torque of abutments varies depending on the type of abutment and implant. Implant Swiss abutments exhibited the highest loosening torque, while SNUCone abutments showed the lowest. The highest average abutment removal torque was observed in the Implant Swiss implant with the SNUCone abutment, and the lowest was in the Dentis implant with the Implant Swiss abutment. The abutment type has a limited effect on this parameter, which may be influenced by other factors such as the overall design of the implant systems.

European Dental Research and Biomaterials JournalVol. 06(01/02)
Isfahan University of Medical Sciences (IR)
Openalex Percentile: Top 83%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.