Noninvasive Repair of Dual Lumen Cannula Crack in a Patient Supported With Extracorporeal Membrane Oxygenation

BackgroundThe use of extended duration extracorporeal membrane oxygenation (ECMO) has significantly increased during the COVID-19 pandemic. However, prolonged ECMO support heightens the risk of severe complications within the circuit, such as fractures of the cannula, complicating clinical management.InterventionA 16-year-old male presented to the emergency department with a Glasgow Coma Score (GCS) of 3 following a rollover motor vehicle accident. Imaging indicated multifocal intracranial hemorrhage and diffuse axonal injury. He was admitted to the intensive care unit, where Neurosurgery placed an intracranial pressure (ICP) monitor and extraventricular drain (EVD). The patient developed hypoxemic respiratory failure due to acute respiratory distress syndrome (ARDS). With parental consent, veno-venous (VV) ECMO support was initiated. Cannulation went smoothly, and ECMO was maintained for nine days with minimal complications. However, on day nine, the cannula site began to ooze for 13 days despite several interventions, including dressing changes and attempts at suture repair. Inspection revealed a crack on the return side of the bicaval cannula, causing the bleeding. To avoid cannula exchange, Dermabond was used to seal the defect.OutcomeFollowing the Dermabond application, the cannula site was hemostatic, and imaging showed stable cannula positioning. The patient remained on ECMO for an additional 49 days, ultimately achieving respiratory recovery and successful decannulation. He was later discharged to an inpatient rehabilitation facility and then home with a good prognosis.ConclusionDermabond application for ECMO circuit repair proves is an effective strategy for minimizing risks related to cannula exchange during circuit dysfunction.

Authors

Institutions

Publication Details

Journal
The American Surgeon
Published
2025-06-02
DOI
https://doi.org/10.1177/00031348251346544
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Noninvasive Repair of Dual Lumen Cannula Crack in a Patient Supported With Extracorporeal Membrane Oxygenation

Lucas P. Neff, Leah M. Sieren, Mitchell R. Ladd, Kristen A. Zeller et al.
The American Surgeon
Mechanical Circulatory Support Devices
article

Noninvasive Repair of Dual Lumen Cannula Crack in a Patient Supported With Extracorporeal Membrane Oxygenation

Lucas P. Neff, Leah M. Sieren, Mitchell R. Ladd, Kristen A. Zeller, Scott C. Copus, John K. Petty, William J. Dugal, Thomas Pranikoff, Heather D. Ots
article en

Abstract

BackgroundThe use of extended duration extracorporeal membrane oxygenation (ECMO) has significantly increased during the COVID-19 pandemic. However, prolonged ECMO support heightens the risk of severe complications within the circuit, such as fractures of the cannula, complicating clinical management.InterventionA 16-year-old male presented to the emergency department with a Glasgow Coma Score (GCS) of 3 following a rollover motor vehicle accident. Imaging indicated multifocal intracranial hemorrhage and diffuse axonal injury. He was admitted to the intensive care unit, where Neurosurgery placed an intracranial pressure (ICP) monitor and extraventricular drain (EVD). The patient developed hypoxemic respiratory failure due to acute respiratory distress syndrome (ARDS). With parental consent, veno-venous (VV) ECMO support was initiated. Cannulation went smoothly, and ECMO was maintained for nine days with minimal complications. However, on day nine, the cannula site began to ooze for 13 days despite several interventions, including dressing changes and attempts at suture repair. Inspection revealed a crack on the return side of the bicaval cannula, causing the bleeding. To avoid cannula exchange, Dermabond was used to seal the defect.OutcomeFollowing the Dermabond application, the cannula site was hemostatic, and imaging showed stable cannula positioning. The patient remained on ECMO for an additional 49 days, ultimately achieving respiratory recovery and successful decannulation. He was later discharged to an inpatient rehabilitation facility and then home with a good prognosis.ConclusionDermabond application for ECMO circuit repair proves is an effective strategy for minimizing risks related to cannula exchange during circuit dysfunction.

The American SurgeonVol. 91(7)
Atrium Health Wake Forest Baptist (US)
Good health and well-being
Openalex Percentile: Top 88%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.