BEGAM, a new apparatus for the identification of β emitter contaminants in radiopharmaceuticals
The BEGAM project aims at developing a portable detector capable of performing γ/β coincidence and anti-coincidence measurements to estimate the presence of long-lived β-contaminants in radiopharmaceutical solutions. A preliminary prototype of the apparatus is under construction. It is composed of four sectors of Caesium Iodide (CsI) scintillators surrounding a cylindrical plastic scintillator. The scintillation readout is performed via silicon photomultiplier (SiPM) and after some preamplification the signals are sent to a commercial digitizer. Geant4 simulations of the detector's response have been developed and validated using gamma and electron sources. Preliminary γ/β coincidence measurements from a 207Bi source have been performed. Near-future plans include testing the BEGAM detector to estimate the activity of 99Mo contaminants in 99mTc solutions.
Authors
- A. Nannini (ORCID: https://orcid.org/0000-0003-0659-7648)
- S. Piantelli (ORCID: https://orcid.org/0000-0002-0669-2787)
- P. Saletti
- G. Casini (ORCID: https://orcid.org/0000-0002-0270-4437)
- M. Perri (ORCID: https://orcid.org/0000-0002-2988-7157)
- Alessandro Passeri
- Giulia Nicastro
Institutions
- Istituto Nazionale di Fisica Nucleare, Sezione di Firenze (IT)
Publication Details
- Journal
- HNPS Advances in Nuclear Physics
- Published
- 2025-05-31
- DOI
- https://doi.org/10.12681/hnpsanp.8134
- Primary Topic
- Nuclear and radioactivity studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00