FULLY AUTOMATED MEASUREMENT OF PAEDIATRIC CEREBRAL PALSY PELVIC RADIOGRAPHS USING MACHINE LEARNING: EXTERNAL VALIDATION USING A NATIONAL SURVEILLANCE DATABASE

The radiographic analysis required for a national cerebral palsy (CP) hip surveillance programme is resource intensive. BoneFinder® is a machine-learning tool that can automatically calculate Reimer's migration percentage (RMP) from pelvic radiographs. HipScreen is a smartphone application that can partially automate RMP measurement. Three RMP measurement methods were compared across the same set of radiographs: 1) routine manual measurements performed by clinical experts from the CP Integrated Pathway Scotland (CPIPS) database, 2) automated measurements using BoneFinder® and 3) measurements performed by two clinicians using HipScreen. 509 AP pelvic radiographs (1,018 hips; mean age:7.4 years) were selected at random from the CPIPS database. GMFCS levels were I (n=69), II (n=37), III (n=97), IV (n=120) and V (n=186). The absolute mean difference in RMP between BoneFinder® and CPIPS measurements, BoneFinder® and HipScreen and CPIPS and HipScreen measurements was 6.3%, 4.6% and 5.2% respectively. Interobserver reliability (ICC) of RMP measurement across the three methods was excellent (ICC = .92, P<.001, 95% CI .90–.93). Good to excellent ICC and correlation were found between BoneFinder® and CPIPS measurements (ICC = .87, P<.001, 95% CI .75–.93, r=.90) and HipScreen and CPIPS measurements (ICC = .91, P<.001, 95% CI .87–.94, r=.93). The area under the receiver operating characteristic curve for BoneFinder®'s and HipScreen's ability to detect a RMP ≥30/≥40% was .96/.98 and .97/.99, respectively. Fully automated RMP measurements were highly reliable with clinically acceptable measurement error. BoneFinder® appears to perform well in analysis of radiographs in CP children who may have challenging radiographic anatomy.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2025-06-02
DOI
https://doi.org/10.1302/1358-992x.2025.4.001
Primary Topic
Medical Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

FULLY AUTOMATED MEASUREMENT OF PAEDIATRIC CEREBRAL PALSY PELVIC RADIOGRAPHS USING MACHINE LEARNING: EXTERNAL VALIDATION USING A NATIONAL SURVEILLANCE DATABASE

K. Hughes, Claudia Lindner, Jinyi Lang, T.F. Cootes et al.
Orthopaedic Proceedings
Medical Imaging and Analysis
article

FULLY AUTOMATED MEASUREMENT OF PAEDIATRIC CEREBRAL PALSY PELVIC RADIOGRAPHS USING MACHINE LEARNING: EXTERNAL VALIDATION USING A NATIONAL SURVEILLANCE DATABASE

K. Hughes, Claudia Lindner, Jinyi Lang, T.F. Cootes, Daniel C. Perry, Mark S. Gaston, Jessenka Luzar
article en

Abstract

The radiographic analysis required for a national cerebral palsy (CP) hip surveillance programme is resource intensive. BoneFinder® is a machine-learning tool that can automatically calculate Reimer's migration percentage (RMP) from pelvic radiographs. HipScreen is a smartphone application that can partially automate RMP measurement. Three RMP measurement methods were compared across the same set of radiographs: 1) routine manual measurements performed by clinical experts from the CP Integrated Pathway Scotland (CPIPS) database, 2) automated measurements using BoneFinder® and 3) measurements performed by two clinicians using HipScreen. 509 AP pelvic radiographs (1,018 hips; mean age:7.4 years) were selected at random from the CPIPS database. GMFCS levels were I (n=69), II (n=37), III (n=97), IV (n=120) and V (n=186). The absolute mean difference in RMP between BoneFinder® and CPIPS measurements, BoneFinder® and HipScreen and CPIPS and HipScreen measurements was 6.3%, 4.6% and 5.2% respectively. Interobserver reliability (ICC) of RMP measurement across the three methods was excellent (ICC = .92, P<.001, 95% CI .90–.93). Good to excellent ICC and correlation were found between BoneFinder® and CPIPS measurements (ICC = .87, P<.001, 95% CI .75–.93, r=.90) and HipScreen and CPIPS measurements (ICC = .91, P<.001, 95% CI .87–.94, r=.93). The area under the receiver operating characteristic curve for BoneFinder®'s and HipScreen's ability to detect a RMP ≥30/≥40% was .96/.98 and .97/.99, respectively. Fully automated RMP measurements were highly reliable with clinically acceptable measurement error. BoneFinder® appears to perform well in analysis of radiographs in CP children who may have challenging radiographic anatomy.

Orthopaedic ProceedingsVol. 107-B(SUPP_4)
Royal Hospital for Children (GB)
Openalex Percentile: Top 88%
Medical Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.