Numerical investigation of blood rheology in ventricular assist devices: effects on performance and shear stress

This study evaluates whether Newtonian models can replace non-Newtonian models in ventricular assist device (VAD) simulations. Five rheological models were compared in an axial-flow VAD using ANSYS CFX. High-shear conditions (92% > 300 s<sup>-1</sup> rendered non-Newtonian effects negligible, with errors <1% for pressure rise, efficiency, and torque. Wall shear stress variations were minimal (±5 Pa) and below hemolysis thresholds. Newtonian models suffice for performance predictions in high-shear regions, reducing computational costs by 30-50%. However, localized non-Newtonian effects in stagnation zones may need analysis for thrombogenicity. These findings streamline VAD design without compromising accuracy.

Authors

Institutions

Publication Details

Journal
Computer Methods in Biomechanics & Biomedical Engineering
Published
2025-06-02
DOI
https://doi.org/10.1080/10255842.2025.2512876
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Numerical investigation of blood rheology in ventricular assist devices: effects on performance and shear stress

Mourad Taha‐Janan, Mohamed Bounouib
Computer Methods in Biomechanics & Biomedical Engineering
Mechanical Circulatory Support Devices
article

Numerical investigation of blood rheology in ventricular assist devices: effects on performance and shear stress

Mourad Taha‐Janan, Mohamed Bounouib
article en

Abstract

This study evaluates whether Newtonian models can replace non-Newtonian models in ventricular assist device (VAD) simulations. Five rheological models were compared in an axial-flow VAD using ANSYS CFX. High-shear conditions (92% > 300 s<sup>-1</sup> rendered non-Newtonian effects negligible, with errors <1% for pressure rise, efficiency, and torque. Wall shear stress variations were minimal (±5 Pa) and below hemolysis thresholds. Newtonian models suffice for performance predictions in high-shear regions, reducing computational costs by 30-50%. However, localized non-Newtonian effects in stagnation zones may need analysis for thrombogenicity. These findings streamline VAD design without compromising accuracy.

Computer Methods in Biomechanics & Biomedical EngineeringVol. 28(11)
Mohammed V University (MA), Arts et Métiers (FR)
Openalex Percentile: Top 88%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Numerical investigation of blood rheology in ventricular assist devices: effects on performance and shear stress — Mourad Taha‐Janan, Mohamed Bounouib · Computer Methods in Biomechanics & Biomedical Engineering (2025) | TGRS Research Map | TGRS