Wearable Device for Continuous and Real-Time Monitoring of Human Sweat Sodium
Wearable sweat-sensing devices hold significant potential for non-invasive, continuous health monitoring. However, challenges such as ensuring data accuracy, sensor reliability, and measurement stability persist. This study presents the development of a wearable system for the real-time monitoring of human sweat sodium levels, addressing these challenges through the integration of a novel microfluidic chip and a compact potentiostat. The microfluidic chip, fabricated using hydrophilic materials and designed with vertical channels, optimizes sweat flow, prevents backflow, and minimizes sample contamination. The developed wearable potentiostat, as a measurement device, precisely measures electrical currents across a wide dynamic range, from nanoamperes to milliamperes. Validation results demonstrated accurate sodium concentration measurements ranging from 10 mM to 200 mM, with a coefficient of variation below 4% and excellent agreement with laboratory instruments (intraclass correlation = 0.998). During physical exercise, the device measured a decrease in sweat sodium levels, from 101 mM to 67 mM over 30 min, reflecting typical physiological responses to sweating. These findings confirm the system's reliability in providing continuous, real-time sweat sodium monitoring. This work advances wearable health-monitoring technologies and lays the groundwork for applications in fitness optimization and personalized hydration strategies. Future work will explore multi-biomarker integration and broader clinical trials to further validate the system's potential.
Authors
- Asrulnizam Abd Manaf (ORCID: https://orcid.org/0000-0002-6198-263X)
- Nur Fatin Adini Ibrahim (ORCID: https://orcid.org/0000-0002-8900-8337)
- Zulkarnay Zakaria (ORCID: https://orcid.org/0000-0003-4490-7336)
- Nurul Izni Rusli (ORCID: https://orcid.org/0000-0001-8815-321X)
- Anas Mohd Noor (ORCID: https://orcid.org/0000-0002-0410-3635)
- M. Mazalan (ORCID: https://orcid.org/0000-0002-4946-2893)
- Muhammad Salman Al Farisi (ORCID: https://orcid.org/0000-0003-4870-9337)
- Asnida Abdul Wahab (ORCID: https://orcid.org/0000-0003-3222-1732)
- Norhayati Sabani
Institutions
- Universiti Sains Malaysia (MY)
- University of Technology Malaysia (MY)
- Hiroshima City University (JP)
- Universiti Malaysia Perlis (MY)
Publication Details
- Journal
- Sensors
- Published
- 2025-05-30
- DOI
- https://doi.org/10.3390/s25113467
- Citations
- 5
- Primary Topic
- Analytical Chemistry and Sensors
- Type
- article
- Field-Weighted Citation Impact
- 3.68