Case of Frozen Leads: An Old but Persistent Challenge

Frozen leads are a rare but clinically relevant complication during cardiac device replacement. A 77-year-old man undergoing pacemaker upgrade had atrial and ventricular leads firmly adhered to the generator header despite set screw removal and ethanol application. As specialized extraction tools were unavailable, an orthopedic nipper and forceps were used as a locking mechanism to dismantle the header, successfully freeing the leads without damage. This is the first reported case in which fully exposed lead pins could not be extracted. While adhesion limited to the pin port may be addressed with ethanol or fine needles, adhesion from the lead base requires header destruction, increasing damage risk. Specialized tools may not always be available, and in such scenarios, the described method may serve as a practical alternative. Thus, this case underscores the critical importance of meticulous blood and fluid removal during cardiac implantable electronic device (CIED) implantation to prevent frozen leads.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2025-06-02
DOI
https://doi.org/10.1111/pace.15218
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Case of Frozen Leads: An Old but Persistent Challenge

Toshiro Shinke, Yoshikazu Suzuki, Hiroto Sugiyama, Yuya Nakamura et al.
Pacing and Clinical Electrophysiology
Cardiac pacing and defibrillation studies
article

Case of Frozen Leads: An Old but Persistent Challenge

Toshiro Shinke, Yoshikazu Suzuki, Hiroto Sugiyama, Yuya Nakamura, Taku Asano, Yoshimi Onishi, Masanori Ochiai, Takamasa Ishikawa, Rimpei Ueno
article en

Abstract

Frozen leads are a rare but clinically relevant complication during cardiac device replacement. A 77-year-old man undergoing pacemaker upgrade had atrial and ventricular leads firmly adhered to the generator header despite set screw removal and ethanol application. As specialized extraction tools were unavailable, an orthopedic nipper and forceps were used as a locking mechanism to dismantle the header, successfully freeing the leads without damage. This is the first reported case in which fully exposed lead pins could not be extracted. While adhesion limited to the pin port may be addressed with ethanol or fine needles, adhesion from the lead base requires header destruction, increasing damage risk. Specialized tools may not always be available, and in such scenarios, the described method may serve as a practical alternative. Thus, this case underscores the critical importance of meticulous blood and fluid removal during cardiac implantable electronic device (CIED) implantation to prevent frozen leads.

Pacing and Clinical ElectrophysiologyVol. 48(8)
SHOWA Medical University (JP)
Openalex Percentile: Top 78%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Case of Frozen Leads: An Old but Persistent Challenge — Toshiro Shinke, Yoshikazu Suzuki, et al. · Pacing and Clinical Electrophysiology (2025) | TGRS Research Map | TGRS