Reduction of radiofrequency induced implant heating via flexible metasurface shielding at 7 T

Passive implanted devices are commonly contraindicated at ultra-high field MRI due to the risk of radiofrequency heating. Mitigation of this risk has come in many forms, such as modifying implant materials or creating novel radiofrequency coils. These methods require substantial involvement from manufacturers and may not benefit patients with existing implants. In this study, a tailored metasurface design is demonstrated to improve implant safety at 7 T by shielding the local B<sub>1</sub><sup>+</sup> field. A prototype metasurface was designed and implemented with a unit cell size of 15 mm using discrete capacitors of 30 pF values. Phantom and human body model simulations were used to validate differences in the SAR distribution with and without the metasurface. Fiber optic temperature probes were used to measure temperature increase across two representative orthopedic screws placed inside a tissue mimicking phantom during a high-SAR sequence. Phantom and in-vivo imaging were performed to assess the metasurface effect on image quality. With the metasurface, an average maximum temperature decrease of 0.50 °C or 34.9 % near the implant was observed. RF field simulations yielded similar decreases in SAR for the phantom (40.7 %) and substantial decreases for the in-vivo leg model (97 %). Phantom image SNR showed a global 8.5 % decrease with the metasurface while in-vivo images showed a 4.8 % decrease in SNR, with the region in its immediate vicinity experiencing substantial signal drop. These results demonstrate the feasibility of a metasurface designed to substantially reduce local RF induced heating with only minor degradation of image quality. Future work will focus on refinement of the metasurface design and further in-vivo testing.

Authors

Institutions

Publication Details

Journal
Journal of Magnetic Resonance
Published
2025-06-02
DOI
https://doi.org/10.1016/j.jmr.2025.107918
Citations
1
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reduction of radiofrequency induced implant heating via flexible metasurface shielding at 7 T

Ravinder Reddy, Wyger M. Brink, Samir Mehta, Anshuman Swain et al.
1 citations
Journal of Magnetic Resonance
Advanced MRI Techniques and Applications
1.36
article

Reduction of radiofrequency induced implant heating via flexible metasurface shielding at 7 T

Ravinder Reddy, Wyger M. Brink, Samir Mehta, Anshuman Swain, Paul S. Jacobs, Mark A. Elliott, Neil Wilson, Neeraj Panchal, Pradnya Narvekar
article en
1 citations

Abstract

Passive implanted devices are commonly contraindicated at ultra-high field MRI due to the risk of radiofrequency heating. Mitigation of this risk has come in many forms, such as modifying implant materials or creating novel radiofrequency coils. These methods require substantial involvement from manufacturers and may not benefit patients with existing implants. In this study, a tailored metasurface design is demonstrated to improve implant safety at 7 T by shielding the local B<sub>1</sub><sup>+</sup> field. A prototype metasurface was designed and implemented with a unit cell size of 15 mm using discrete capacitors of 30 pF values. Phantom and human body model simulations were used to validate differences in the SAR distribution with and without the metasurface. Fiber optic temperature probes were used to measure temperature increase across two representative orthopedic screws placed inside a tissue mimicking phantom during a high-SAR sequence. Phantom and in-vivo imaging were performed to assess the metasurface effect on image quality. With the metasurface, an average maximum temperature decrease of 0.50 °C or 34.9 % near the implant was observed. RF field simulations yielded similar decreases in SAR for the phantom (40.7 %) and substantial decreases for the in-vivo leg model (97 %). Phantom image SNR showed a global 8.5 % decrease with the metasurface while in-vivo images showed a 4.8 % decrease in SNR, with the region in its immediate vicinity experiencing substantial signal drop. These results demonstrate the feasibility of a metasurface designed to substantially reduce local RF induced heating with only minor degradation of image quality. Future work will focus on refinement of the metasurface design and further in-vivo testing.

Journal of Magnetic ResonanceVol. 377
National Medical Research Center of Dentistry and Maxillofacial Surgery (RU), University of Pennsylvania (US), University of Twente (NL)
University of Pennsylvania, Leibniz-Zentrum für Marine Tropenforschung, National Institutes of Health, National Institute on Aging, National Institute of Biomedical Imaging and Bioengineering
Openalex Percentile: Top 19%
Advanced MRI Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.