Comparison of image quality between photon-counting detector CT and energy-integrating detector coronary CT angiography in heart transplant patients

This study aims to compare the image quality between photon-counting detector computed tomography (PCD-CT) and energy-integrating CT (EID-CT) for coronary artery visualization post-heart transplant (HTx) patients, using a paired comparison within one patient. Consecutive HTx patients who underwent both 3rd generation dual source EID-CT and ultra-high-resolution PCD-CT scans as part of routine clinical care, with a maximum interval of two years between scans were included. Image quality was assessed subjectively using a 5-point Likert scale and quantitatively with the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Coronary segments were scored for presence, stenosis degree, presence of artifacts, and plaque composition. Thirty patients were included (mean age 50 ± 17 years; 21 (70%) men). The median interval between scans was 17[12-24] months. The median calcium score was 9[0-51] on the EID-CT scan and 25[0-92] on the PCD-CT. On the PCD-CT, 90% of coronary segments were rated very good or excellent compared to 71% on EID-CT (p < 0.001). Fewer segments were absent or non-evaluable due to size with PCD-CT compared to EID-CT (30% vs. 34%; p < 0.001). PCD-CT identified more mixed plaques (11% vs. 3%, p < 0.001). SNR and CNR were significantly higher on PCD-CT (22.1 ± 6.1 and 27.8 ± 6.8, p < 0.001) than on EID-CT (14.3 ± 3.4 and 16.9 ± 3.6, p < 0.001). The CT dose index volume was 6 [5-10] mGy on EID-CT and 19 [14-21] mGy on PCD-CT (p < 0.001). Ultra-high-resolution PCD-CT improves image quality compared to EID-CT in HTx patients. With superior SNR and CNR, and fewer non-evaluable segments, PCD-CT offers enhanced coronary artery visualization, though at the cost of an increased radiation dose.

Authors

Institutions

Publication Details

Journal
The International Journal of Cardiovascular Imaging
Published
2025-06-02
DOI
https://doi.org/10.1007/s10554-025-03433-7
Citations
1
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of image quality between photon-counting detector CT and energy-integrating detector coronary CT angiography in heart transplant patients

Marcel L. Dijkshoorn, Ricardo P.J. Budde, Olivier C. Manintveld, Simran P. Sharma et al.
1 citations
The International Journal of Cardiovascular Imaging
Advanced X-ray and CT Imaging
0.67
article

Comparison of image quality between photon-counting detector CT and energy-integrating detector coronary CT angiography in heart transplant patients

Marcel L. Dijkshoorn, Ricardo P.J. Budde, Olivier C. Manintveld, Simran P. Sharma, J.P. Groen, Alexander Hirsch
article en
1 citations

Abstract

This study aims to compare the image quality between photon-counting detector computed tomography (PCD-CT) and energy-integrating CT (EID-CT) for coronary artery visualization post-heart transplant (HTx) patients, using a paired comparison within one patient. Consecutive HTx patients who underwent both 3rd generation dual source EID-CT and ultra-high-resolution PCD-CT scans as part of routine clinical care, with a maximum interval of two years between scans were included. Image quality was assessed subjectively using a 5-point Likert scale and quantitatively with the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Coronary segments were scored for presence, stenosis degree, presence of artifacts, and plaque composition. Thirty patients were included (mean age 50 ± 17 years; 21 (70%) men). The median interval between scans was 17[12-24] months. The median calcium score was 9[0-51] on the EID-CT scan and 25[0-92] on the PCD-CT. On the PCD-CT, 90% of coronary segments were rated very good or excellent compared to 71% on EID-CT (p < 0.001). Fewer segments were absent or non-evaluable due to size with PCD-CT compared to EID-CT (30% vs. 34%; p < 0.001). PCD-CT identified more mixed plaques (11% vs. 3%, p < 0.001). SNR and CNR were significantly higher on PCD-CT (22.1 ± 6.1 and 27.8 ± 6.8, p < 0.001) than on EID-CT (14.3 ± 3.4 and 16.9 ± 3.6, p < 0.001). The CT dose index volume was 6 [5-10] mGy on EID-CT and 19 [14-21] mGy on PCD-CT (p < 0.001). Ultra-high-resolution PCD-CT improves image quality compared to EID-CT in HTx patients. With superior SNR and CNR, and fewer non-evaluable segments, PCD-CT offers enhanced coronary artery visualization, though at the cost of an increased radiation dose.

The International Journal of Cardiovascular ImagingVol. 42(3)
Erasmus MC (NL), Erasmus MC Cancer Institute (NL), Erasmus University Rotterdam (NL)
Openalex Percentile: Top 31%
Advanced X-ray and CT Imaging
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.