Abstract Background The mechanical performance assessment of orthopedic fixation systems is computationally and experimentally challenging due to the complex geometrical and mechanical features of bones. Non-contact experimental techniques, widely adopted in several engineering fields, is shown to overcome these issues. Objective This work discusses a comparative experimental investigation into specimens mimicking healthy humerus bones and fractured bones subject to an innovative surgery procedure and to a classical technique referred to as the gold standard surgery. The new surgery consists in the installation of an external fixation mechanism that constrains, according to different spatial patterns, a certain number of titanium slender bars inserted and clamped into the fractured bones. Methods The mechanical properties of artificial bones are characterized through compressive tests, while the morphology of the fracture surface is analyzed using a scanning electron microscope. A three-dimensional laser vibrometer is used to measure the resonance frequencies, mode shapes, damping ratios, and mechanical waves propagating from the actuators across the surface of the bones. Results The results provide insights into which configuration of the fixator performs better for a fast recovery. Based on the observed dynamic behaviors, the optimal configuration of the fixator offers performance that is comparable to, or potentially better than, the gold standard surgical procedure. Conclusions The novelty and feasibility of the adopted experimental approach paves the way towards the adoption of advanced non-contact techniques for the mechanical characterization of complex, non-homegenous and anisotropic materials and structures in biomedical applications enabling also data-driven models of the systems.

Authors

Institutions

Publication Details

Journal
Experimental Mechanics
Published
2025-06-02
DOI
https://doi.org/10.1007/s11340-025-01198-7
Citations
2
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
4.01

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Exploring Humerus Bone’s Fracture Patterns and Fixation Systems Via Laser Vibrometry

Stefano Gumina, Walter Lacarbonara, Vittorio Candela, Claudia Sergi et al.
2 citations
Experimental Mechanics
Shoulder Injury and Treatment
4.01
article

Exploring Humerus Bone’s Fracture Patterns and Fixation Systems Via Laser Vibrometry

Stefano Gumina, Walter Lacarbonara, Vittorio Candela, Claudia Sergi, Biagio Carboni, T. Valente, Jacopo Tirillò, Sawan Kumar Guruva
article en
2 citations

Abstract

Abstract Background The mechanical performance assessment of orthopedic fixation systems is computationally and experimentally challenging due to the complex geometrical and mechanical features of bones. Non-contact experimental techniques, widely adopted in several engineering fields, is shown to overcome these issues. Objective This work discusses a comparative experimental investigation into specimens mimicking healthy humerus bones and fractured bones subject to an innovative surgery procedure and to a classical technique referred to as the gold standard surgery. The new surgery consists in the installation of an external fixation mechanism that constrains, according to different spatial patterns, a certain number of titanium slender bars inserted and clamped into the fractured bones. Methods The mechanical properties of artificial bones are characterized through compressive tests, while the morphology of the fracture surface is analyzed using a scanning electron microscope. A three-dimensional laser vibrometer is used to measure the resonance frequencies, mode shapes, damping ratios, and mechanical waves propagating from the actuators across the surface of the bones. Results The results provide insights into which configuration of the fixator performs better for a fast recovery. Based on the observed dynamic behaviors, the optimal configuration of the fixator offers performance that is comparable to, or potentially better than, the gold standard surgical procedure. Conclusions The novelty and feasibility of the adopted experimental approach paves the way towards the adoption of advanced non-contact techniques for the mechanical characterization of complex, non-homegenous and anisotropic materials and structures in biomedical applications enabling also data-driven models of the systems.

Experimental MechanicsVol. 65(7)
Sapienza University of Rome (IT)
Sapienza Università di Roma
Openalex Percentile: Top 7%
Shoulder Injury and Treatment
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.